128
English
12. Technical Data
OXYMAT 3
Product category according to
93/42/EEC
IIa
Dimensions WxHxD in mm 400 x 700 x 350
Weight approx. 20 kg
Temperature range
– Operation
– Storage
+10 to +40 °C
- 20 to +70 °C
Power supply 230V, 50 Hz
Power output / current consumption 360 W / 1.6 A
Mains fuse EN 60127-2 T 2.5 A H 250V
Classification acc. to EN 60601-1
– Protection against electric shock
– Degree of protection against
electric shock
Class 2
Type B
Electromagnetic compatibility
– Radio interference suppression
– Radio interference resistance
EN 55011 (VDE
0875 Pt11)
IEC 1000-4 Parts
2-6 and 11
Sound pressure level
40 dB(A)
(from serial num-
ber 20.000)
O
2
output (values after 10 min.
operation at + 20 °C,
50% relative humidity and
1013 mbar)
95 +1/-3 vol% O
2
at 1–4 l/min
90 ±3 vol% O
2
at
5 l/min
O
2
output at 2000 m above msl
90 ±3 vol% O
2
at
5 l/min
Oxygen outlet pressure 60 kpa
Flow rate (read at top of ball)
0.5 – 5.5 l/min
0.2 – 3.0 l/min
OSCI O
2
status indicator as per
EN ISO 8359
Warning at < 82 % by vol. of O
2
present
Fault warning Visual and acoustic
Actuating pressure of pressure
relief mechanism 300 kPa
Servicing
at least once a
year or every
5000 operating
hours
Subject to technical change without notice.