EasyManua.ls Logo

Welch Allyn Connex VSM 6000 series - NIBP and SpO2;SpHb Messages

Welch Allyn Connex VSM 6000 series
266 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
the Port entry field and enter the port number. The range
of entry is 0 to 65535.
Note The DNS name option is only available in the
following conditions:
The version of the installed radio is 3.00.02 or greater
The radio is disabled
No radio is installed
DHCP Enable the device to connect to a Network Rendezvous
Service (NRS) by entering a port number and then
connecting at an address provided by the DHCP43
response. Touch in the Port entry field and enter the
port number. The range of entry is 0 to 65535. After
touching Test and successfully connecting to the
server, the NRS IP addresses appear onscreen.
Restore defaults Touch Restore defaults to restore settings for the
selected option to the default values.
Test Touch Test to test the connection to the configured
server.
6. Do one of the following:
To continue in the Advanced Settings, touch another tab.
To exit the Advanced Settings and return to the Home tab, touch Exit.
Service
The Service tab presents numerous settings and controls typically accessed by
authorized service or biomedical engineering personnel to configure, maintain, test, and
update the device. For example, the Service tab enables authorized users to save device
configurations to a USB flash drive and then load saved configurations to other devices.
Systems and devices configured with the PartnerConnect™ service feature also have
access to remote diagnostics, troubleshooting, and software upgrades.
For a description of service-related advanced settings, see the service manual for this
product.
Directions for use Advanced settings 205

Table of Contents

Other manuals for Welch Allyn Connex VSM 6000 series

Related product manuals