Protection classifications, all monitor configurations
Weight (including battery) Standard chassis: 9.5 lb (4.3 kg)
Extended chassis: 10.4 lb (4.7 kg)
Integrated Wall System
Height 10.5 in. (26.8 cm)
Width 39.9 in. (101.4 cm)
Depth 7.5 in. (19.1 cm)
Weight (including battery) 14.1 lb. (6 kg)
Graphical display resolution
Display area 8 in. (H) x 4 in. (V) (19.5 [H] cm x 11.3 [V] cm)
Pixels 1024 (H) x 600 (V)
Pixel arrangement RGB (red, green, blue)
Color depth 16 bits per pixel
Speaker volume
Vital Signs Monitor 6000 Series
High priority alarm sound pressure level 47-92 dBA at 1.0 meter
Medium priority alarm sound pressure
level
45-82 dBA at 1.0 meter
Integrated Wall System
High priority alarm sound pressure level 42-78 dBA at 1.0 meter
Medium priority alarm sound pressure
level
42-75 dBA at 1.0 meter
Alarm and pulse tones
per IEC 60601-1-8
Pulse frequency (f
0
) 150 – 1000 Hz
Number of harmonic components in the
range 300 Hz to 4000 Hz
minimum of 4
Effective pulse duration (t
d
) high priority: 75 –200 ms
medium and low priority: 125 – 250 ms
Rise time (t
r
) 10 – 40% of t
d
Fall time
a
(t
f
)
t
f
< t
s
– t
r
222 Specifications Welch Allyn Connex
®
Devices