Alarm limits Upper limit range of entry Lower limit range of entry
RRa
Adult: 5 to 150 breaths per minute. Factory
default: 50 breaths per minute.
Adult: 0 to 145 breaths per minute. Factory
default: 3 breaths per minute.
Patient movement Exit sensitivity: 1 to 6. Factory default: 3
EH > 1 minute
Off
Alarm delays
Physiological parameter Default delay
(seconds)
Configurable
range
(seconds)
1
Sum of default
delays
(seconds)
2
Sum of
configurable
delays
(seconds)
Pulse rate
SpO2 3 3–120 4 4–121
Motion (EarlySense) 3 3–120 4 4–121
SpO2
10 0–30 11 1–31
SpHb
10 0–10 11 1–11
etCO2
15 0–30 16 1–31
etCO2 no breath detected
Adult: 30 10–60 31 11–61
Pediatric: 20 10–60 21 11–61
Neonate: 15 10–60 16 11–61
Motion low confidence RR/PR
alarm delay
180 180–1800 181 181–1801
Respiration rate
RRa (Masimo) 10 0–60 11 1–61
CO2 10 0–60 11 1–61
Motion (EarlySense) 180 10–180 181 11–181
1
These delays are configurable in Advanced settings.
2
When the device determines the alarm condition, it generates alarm signals (visual and auditory) and notifies any
system to which it is connected within one second.
Note The maximum alarm signal generation delay of a technical alarm when connected to a central station
is 4 seconds.
Note There are no inherent alarm condition delays in the system for temperature or blood pressure alarms.
Consider ALL of these alarm delays when setting up your monitor and remote alarm system.
Directions for use Specifications 235