Remote Operation
SCPI Commands for Digital Interfaces
132 Operating Manual for XDC Series Power Supply
Configure
Other
Protection
Mechanisms
Fold Protection
Fold protection causes the supply to shut down if the selected regulation mode is
entered. A delay time may be specified as well.
To set the fold mode:
OUTP:PROT:FOLD {CC|CV|CP|NONE}
Where:
NONE indicates fold protection is disabled.
CC indicates the supply will shut down due to constant current condition.
CV indicates the supply will shut down due to constant voltage.
CP indicates the supply will shut down due to constant power.
The default value is NONE.
To set the fold delay:
OUTP:PROT:FOLD:DEL <delay_time>
Where:
<delay_time> is a value in the range of 0–60 seconds.
Units may be specified as “ms” (milliseconds) “s” (seconds), or “min” (minutes). If
no units are specified, the default seconds are assumed. (Increments of 0.1s are
allowed.) The default value is 0.5 second.