Remote Operation
Remote Analog Operation
112 Operating Manual for XDC Series Power Supply
Using
Remote
Analog
Control
Connect your programming voltage sources and monitors, ensuring that the
appropriate 0– 5V or 0–10V range has been configured and selected. (See page 75.)
With the Analog Control mode selected, varying the programming source from 0 to
5V (or 10) will vary the output voltage or current from 0 to its rated maximum. A
voltage in the range of 0–5V (or 10) on the corresponding monitor line is
proportional to 0–100% of the rated output.
The output V and I readback lines are active for all operating modes as are the user
interface lines. The analog programming inputs are active only when selected using
the procedure in “Select Remote Control Source” on page 75. You can select analog
remote control or either the output voltage or current, or both. When only voltage or
current is being remotely controlled, the other setpoint is under front panel local
control.