Calibration
Entering Calibration Mode
180 Operating Manual for XDC Series Power Supply
Front Panel To access calibration mode via the front panel:
If your password code is incorrect, the prompt displays Incorrect code and
automatically returns to Code ####. Either try again or press EXIT to escape.
If your password code is correct, the prompt displays Output V Cal. This is the
first of 12 available options. The 12 sets of parameters that must be adjusted during
calibration are:
Output V Cal: Calibrate voltage output and readback.
Output I Cal: Calibrate current output and readback.
ANLG V PGM 5V: Calibrate 5V analog programming interface for setting
voltage output.
ANLG V RB 5V: Calibrate 5V analog programming interface for monitoring
voltage output.
ANLG I PGM 5V: Calibrate 5V analog programming interface for setting
current output.
ANLG I RB 5V: Calibrate 5V analog programming interface for monitoring
current output.
ANLG V PGM 10V: Calibrate 10V analog programming interface for setting
voltage output.
ANLG V RB 10V: Calibrate 10V analog programming interface for monitoring
voltage output.
ANLG I PGM 10V: Calibrate 10V analog programming interface for setting
current output.
ANLG I RB 10V: Calibrate 10V analog programming interface for monitoring
current output.
Factory Cal: Lets you restore the factory calibration constants.
Change Code: Lets you change the password code.
Step # Do This You Will See
1 ERROR MSGS
2
Press 2 times.
CALIBRATION
3
Code #####
4 Enter the calibration security code.
The factory code is “0000”.
Code 0000
5 You are now in the
Calibration menu.
Output V Cal
MENU
ENTER
ENTER