EasyManua.ls Logo

Xo FLEX - Page 51

Xo FLEX
95 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XO FLEX – Instructions for use Configuration
51
Parameter
Display text
Factory default
configuration
Possible settings
Amount of spray water –
Level 3
W LEVEL 3
50 ml/min
Peristaltic pump i
rrigation
flow rate: 10 – 90 ml/min ±
20%
Amount of spray water –
Level 2
W LEVEL 2
30 ml/min
Peristaltic pump i
rrigation
flow rate: 10 – 90 ml/min ±
20%
Amount of spray water –
Level 1
W LEVEL 1
10 ml/min
Peristaltic pump i
rrigation
flow rate: 10 – 90 ml/min ±
20%
Turbine (Air instrument)
Drive air – Level 3
LEVEL 3
100%
0 – 100%
Drive air – Level 2
LEVEL 2
90%
0 – 100%
Drive air – Level 1
LEVEL 1
70%
0 – 100%
Amount of spray water –
Level 3
W LEVEL 3
Without turbine
hand-piece (what
you see on the
display): 100 ml/min
With turbine hand-
piece:
approximately 70
ml/min
Without turbine hand-piece:
10 – 100 + 20% ml/min
Amount of spray water –
Level 2
W LEVEL 2
Without turbine
hand-piece (what
you see on the
display): 50 ml/min
With turbine hand-
piece:
approximately 50
ml/min
Without turbine hand-piece:
10 – 100 + 20% ml/min
Amount of spray water –
Level 1
W LEVEL 1
Without turbine
hand-piece (what
you see on the
display): 35 ml/min
With turbine hand-
piece:
approximately 35
ml/min
Without turbine hand-piece:
10 – 100 + 20% ml/min
Amount of spray air
AIR
70%
0 – 100%
Automatic chip blow
CHIP
Enabled
Enabled, Disabled
XO ODONTOSON 360 or other ultrasonic scaler
Power – Level 3
LEVEL 3
100%
0 – 100%
Power – Level 2
LEVEL 2
70%
0 – 100%
Power – Level 1
LEVEL 1
40%
0 – 100%
Amount of irrigation
water – Level 3
W LEVEL 3
40 ml/min
10 – 90 + 10% ml/min
Amount of irrigation
water – Level 2
W LEVEL 2
30 ml/min
10 – 90 + 10% ml/min

Table of Contents