5-8
IM 04P01B01-17E
Measured/computed Data
The FD (FD0) or FY (FYInst, FYTlog1, or FYTlog2) command is used to output the
data.
Syntax
The measured/computed data are output in the following syntax along with the date
and time information for each channel
EACRLF
DATE_yy/mo/ddCRLF
TIME_hh:mi:ss.mmmt_S
1
S
2
S
3
S
4
S
5
S
6
CRLF
s_kcca
1
a
2
a
3
a
4
uuuuuuf
1
dddddEf
2
ppCRLF
••••••••••••••••••••••••••••
ENCRLF
yy Year (00 to 99)
mo Month (01 to 12)
dd Day (01 to 31)
hh Hour (00 to 23)
mi Minute (00 to 59)
ss Second (00 to 59)
mmm Millisecond (000 to 999. A period is placed between seconds
and milliseconds.)
t ‘S’(=summer) or ‘ ’ space(=winter)
S
1
S
2
S
3
S
4
S
5
S
6
Data status
Takes on the values below for the FYTlog1 and FYTlog2
commands. For all other cases, they are all spaces.
S
1
Time change during TLOG computation: T (occurred) or space
(not occurred)
S
2
Power OFF and ON during TLOG computation: P (occurred) or
space (not occurred)
S
3
Data reset during the TLOG computation: R (occurred) or
space (not occurred)
S
4
S
5
S
6
All spaces
s Channel data status (N, D, S, O, B, E)
N: Normal
D: Differential input
S: Skip
O: Over
B: Burnout
E: Error
k Channel type
0: Measurement channel
A: Computation channel
cc Channel number
µR10000: 01 to 06, 0A to 0P
µR20000: 01 to 24, 0A to 1P
a
1
a
2
a
3
a
4
a
1
Alarm status (level 1)
a
2
Alarm status (level 2)
a
3
Alarm status (level 3)
a
4
Alarm status (level 4)
5.2 Output Format of ASCII Data