EasyManuals Logo

YOKOGAWA DL9040 User Manual

YOKOGAWA DL9040
256 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #159 background imageLoading...
Page #159 background image
5-125
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:MEASure:THReshold:TRACe<x>:
LHYSteresis:LEVel
Function Sets the level of the threshold level or queries the
current setting.
Syntax
:MEASure:THReshold:TRACe<x>:
LHYSteresis:LEVel {<NRf>|<Voltage>|<Cur
rent>}
:MEASure:THReshold:TRACe<x>:
LHYSteresis:LEVel?
<x> = 1 to 8
<NRf>, <V
oltage>, and <Current> = See the DL9000
User’s Manual.
Example
:MEASURE:THRESHOLD:TRACE1:LHYSTERESIS:
LEVEL 1
:MEASURE:THRESHOLD:TRACE1:LHYSTERESIS:
LEVEL? -> :MEASURE:THRESHOLD:TRACE1:
LHYSTERESIS:LEVEL 1.000E+00
Description This command is valid when the <Parameter> of
:MEASure:TRACe<x>:AREA<x>:<Parameter>
is
not RISE or FALL.
:MEASure:THReshold:TRACe<x>:MODE
Function Sets the setup mode of the threshold level or queries
the current setting.
Syntax
:MEASure:THReshold:TRACe<x>:MODE {AUTO|
LHYSteresis|ULOWer}
:MEASure:THReshold:TRACe<x>:MODE?
<x> = 1 to 8
Example
:MEASURE:THRESHOLD:TRACE1:
MODE LHYSTERESIS
:MEASURE:THRESHOLD:TRACE1:MODE?
-> :MEASURE:THRESHOLD:TRACE1:
MODE LHYSTERESIS
Description This command is valid when the <Parameter> of
:MEASure:TRACe<x>:AREA<x>:<Parameter>
is
not RISE or FALL.
:MEASure:THReshold:TRACe<x>:ULOWer?
Function Queries all settings related to the upper and lower
limits of the threshold level.
Syntax
:MEASure:THReshold:TRACe<x>:ULOWer?
<x> = 1 to 8
Example
:MEASURE:THRESHOLD:TRACE1:ULOWER?
-> :MEASURE:THRESHOLD:TRACE1:ULOWER:
RANGE 2.000E+00,1.000E+00
Description This command is valid when the <Parameter> of
:MEASure:TRACe<x>:AREA<x>:<Parameter>
is
not RISE or FALL.
:MEASure:THReshold:TRACe<x>:ULOWer:
RANGe
Function Sets the upper and lower limits of the threshold level
or queries the current setting.
Syntax
:MEASure:THReshold:TRACe<x>:ULOWer:
RANGe {(<NRf>,<NRf>)|
(<Voltage>,<Voltage>)|
(<Current>,<Current>)}
:MEASure:THReshold:TRACe<x>:ULOWer:
RANGe?
<x> = 1 to 8
<NRf>, <V
oltage>, and <Current> = See the DL9000
User’s Manual.
Example
:MEASURE:THRESHOLD:TRACE1:ULOWER:
RANGE 1,2
:MEASURE:THRESHOLD:TRACE1:ULOWER:RANGE?
-> :MEASURE:THRESHOLD:TRACE1:ULOWER:
RANGE 2.000E+00,1.000E+00
Description This command is valid when the <Parameter> of
:MEASure:TRACe<x>:AREA<x>:<Parameter>
is
not RISE or FALL.
5.18 MEASure Group

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DL9040 and is the answer not in the manual?

YOKOGAWA DL9040 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDL9040
CategoryTest Equipment
LanguageEnglish

Related product manuals