5-129
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:MEASure:TRACe<x>:AREA<x>:DELay:
REFerence:COUNt
Function Sets the edge detection count of the reference
waveform of the delay measurement between
channels or queries the current setting.
Syntax
:MEASure:TRACe<x>:AREA<x>:DELay:
REFerence:COUNt {<NRf>}
:MEASure:TRACe<x>:AREA<x>:DELay:
REFerence:COUNt?
<x> of TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
<NRf> = 1 to 10
Example
:MEASURE:TRACE1:AREA1:DELAY:REFERENCE:
COUNT 1
:MEASURE:TRACE1:AREA1:DELAY:REFERENCE:
COUNT? -> :MEASURE:TRACE1:AREA1:DELAY:
REFERENCE:COUNT 1
:MEASure:TRACe<x>:AREA<x>:DELay:
REFerence:POLarity
Function Sets the polarity of the reference waveform of the
delay measurement between channels or queries the
current setting.
Syntax
:MEASure:TRACe<x>:AREA<x>:DELay:
REFerence:POLarity {FALL|RISE}
:MEASure:TRACe<x>:AREA<x>:DELay:
REFerence:POLarity?
<x> of TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
Example
:MEASURE:TRACE1:AREA1:DELAY:REFERENCE:
POLARITY FALL
:MEASURE:TRACE1:AREA1:DELAY:REFERENCE:
POLARITY? -> :MEASURE:TRACE1:AREA1:
DELAY:REFERENCE:POLARITY FALL
:MEASure:TRACe<x>:AREA<x>:DELay:
REFerence:TRACe
Function Sets the trace of the reference waveform of the delay
measurement between channels or queries the
current setting.
Syntax
:MEASure:TRACe<x>:AREA<x>:DELay:
REFerence:TRACe {<NRf>}
:MEASure:TRACe<x>:AREA<x>:DELay:
REFerence:TRACe?
<x> of TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
<NRf> = 1 to 8
Example
:MEASURE:TRACE1:AREA1:DELAY:REFERENCE:
TRACE 1
:MEASURE:TRACE1:AREA1:DELAY:REFERENCE:
TRACE? -> :MEASURE:TRACE1:AREA1:DELAY:
REFERENCE:TRACE 1
:MEASure:TRACe<x>:AREA<x>:DELay:
SOURce
Function Sets the reference of the delay measurement
between channels or queries the current setting.
Syntax
:MEASure:TRACe<x>:AREA<x>:DELay:SOURce
{TRACe|TRIGger}
:MEASure:TRACe<x>:AREA<x>:DELay:SOURce?
<x> of TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
Example
:MEASURE:TRACE1:AREA1:DELAY:
SOURCE TRACE
:MEASURE:TRACE1:AREA1:DELAY:SOURCE?
-> :MEASURE:TRACE1:AREA1:DELAY:
SOURCE TRACE
:MEASure:TRACe<x>:AREA<x>:DPRoximal?
Function Queries all settings related to the distal and proximal
values.
Syntax
:MEASure:TRACe<x>:AREA<x>:DPRoximal?
<x> of TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
Example
:MEASURE:TRACE1:AREA1:DPROXIMAL?
-> ;MEASURE:TRACE1:AREA1:DPROXIMAL:
MODE PERCENT;PERCENT 10,20;UNIT
1.000E+00,1.000E+00
Description This command is valid when the <Parameter> of
:MEASure:TRACe<x>:AREA<x>:<Parameter>
is
RISE or FALL.
:MEASure:TRACe<x>:AREA<x>:DPRoximal:
MODE
Function Sets the unit of the distal and proximal values or
queries the current setting.
Syntax
:MEASure:TRACe<x>:AREA<x>:DPRoximal:
MODE {PERCent|UNIT}
:MEASure:TRACe<x>:AREA<x>:DPRoximal:
MODE?
<x> of TRACe<x> = 1 to 8
<x> of AREA<x> = 1 or 2
Example
:MEASURE:TRACE1:AREA1:DPROXIMAL:
MODE PERCENT
:MEASURE:TRACE1:AREA1:DPROXIMAL:MODE?
-> :MEASURE:TRACE1:AREA1:DPROXIMAL:
MODE PERCENT
Description This command is valid when the <Parameter> of
:MEASure:TRACe<x>:AREA<x>:<Parameter>
is
RISE or FALL.
5.18 MEASure Group