Loading...
5-173
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:BYTE
Function Sets the number of bytes of the data of the SPI bus
trigger or queries the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:BYTE {<NRf>}
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:BYTE?
<x> of EVENt<x> = 1 or 2
<x> of DA
TA<x> = 1 or 2
<NRf> = 1 to 4
Example
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
BYTE 1
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
BYTE? -> :TRIGGER:EINTERVAL:EVENT1:
SPIBUS:DATA1:BYTE 1
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:CONDition
Function Sets the determination method (match or not match)
of the data of the SPI bus trigger or queries the
current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:CONDition {FALSe|TRUE}
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:CONDition?
<x> of EVENt<x> = 1 or 2
<x> of DA
TA<x> = 1 or 2
Example
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
CONDITION TRUE
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
CONDITION? -> :TRIGGER:EINTERVAL:
EVENT1:SPIBUS:DATA1:CONDITION TRUE
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:DPOSition
Function Sets the pattern comparison start position of the data
of the SPI bus trigger or queries the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:DPOSition {<NRf>}
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:DPOSition?
<x> of EVENt<x> = 1 or 2
<x> of DA
TA<x> = 1 or 2
<NRf> = 0 to 9999
Example
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
DPOSITION 1
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
DPOSITION? -> :TRIGGER:EINTERVAL:
EVENT1:SPIBUS:DATA1:DPOSITION 1
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:HEXA<x>
Function Sets the data of the SPI bus trigger in hexadecimal
notation.
Syntax
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:HEXA<x> {<String>}
<x> of EVENt<x> = 1 or 2
<x> of DA
TA<x> = 1 or 2
<x> of HEXA<x> = 1 to 4
<String> = 2 characters by combining ‘0’ to ‘F and ‘X’
Example
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
HEXA1 "AB"
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:PATTern<x>
Function Sets the data of the SPI bus trigger in binary notation
or queries the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:PATTern<x> {<String>}
:TRIGger:EINTerval:EVENt<x>:
SPIBus:
DATA<x>:PATTern<x>?
<x> of EVENt<x> = 1 or 2
<x> of DA
TA<x> = 1 or 2
<x> of PATTern<x> = 1 to 4
<String> = 8 characters by combining ‘0,’ ‘1,’ and ‘X’
Example
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
PATTERN1 "10101011"
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
PATTERN1? -> :TRIGGER:EINTERVAL:EVENT1:
SPIBUS:DATA1:PATTERN1 "10101011"
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:SOURce
Function Sets the trace of the data of the SPI bus trigger or
queries the current setting.
Syntax
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:SOURce {<NRf>}
:TRIGger:EINTerval:EVENt<x>:SPIBus:
DATA<x>:SOURce?
<x> of EVENt<x> = 1 or 2
<x> of DA
TA<x> = 1 or 2
<NRf> = 1 to 4
Example
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
SOURCE 1
:TRIGGER:EINTERVAL:EVENT1:SPIBUS:DATA1:
SOURCE? -> :TRIGGER:EINTERVAL:EVENT1:
SPIBUS:DATA1:SOURCE 1
5.29 TRIGger Group