EasyManua.ls Logo

YOKOGAWA DL9040 - Page 227

YOKOGAWA DL9040
256 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5-193
IM 701310-17E
Commands
1
2
3
4
5
6
App
Index
:TRIGger:ENHanced:SPIBus:DATA<x>:
HEXA<x>
Function Sets the data of the SPI bus trigger in hexadecimal
notation.
Syntax
:TRIGger:ENHanced:SPIBus:DATA<x>:
HEXA<x> {<String>}
<x> of DAT
A<x> = 1 or 2
<x> of HEXA<x> = 1 to 4
<String> = 2 characters by combining ‘0’ to ‘F’ and ‘X’
Example
:TRIGGER:ENHANCED:SPIBUS:DATA1:
HEXA1 "AB"
:TRIGger:ENHanced:SPIBus:DATA<x>:
PATTern<x>
Function Sets the data of the SPI bus trigger in binary notation
or queries the current setting.
Syntax
:TRIGger:ENHanced:SPIBus:DATA<x>:
PATTern<x> {<String>}
:TRIGger:ENHanced:SPIBus:DATA<x>:
PATTern<x>?
<x> of DAT
A<x> = 1 or 2
<x> of PATTern<x> = 1 to 4
<String> = 8 characters by combining ‘0,’ ‘1,’ and ‘X’
Example
:TRIGGER:ENHANCED:SPIBUS:DATA1:
PATTERN1 "10101011"
:TRIGGER:ENHANCED:SPIBUS:DATA1:
PATTERN1? -> :TRIGGER:ENHANCED:SPIBUS:
DATA1:PATTERN1 "10101011"
:TRIGger:ENHanced:SPIBus:DATA<x>:
SOURce
Function Sets the trace of the data of the SPI bus trigger or
queries the current setting.
Syntax
:TRIGger:ENHanced:SPIBus:DATA<x>:
SOURce {<NRf>}
:TRIGger:ENHanced:SPIBus:DATA<x>:
SOURce?
<x> = 1 or 2
<NRf> = 1 to 4
Example
:TRIGGER:ENHANCED:SPIBUS:DATA1:SOURCE 1
:TRIGGER:ENHANCED:SPIBUS:DATA1:SOURCE?
-> :TRIGGER:ENHANCED:SPIBUS:DATA1:
SOURCE 1
:TRIGger:ENHanced:SPIBus:MODE
Function Sets the wiring system of the SPI bus trigger (three-
wire or four-wire) or queries the current setting.
Syntax
:TRIGger:ENHanced:SPIBus:MODE {WIRE3|
WIRE4}
:TRIGger:ENHanced:SPIBus:MODE?
Example
:TRIGGER:ENHANCED:SPIBUS:MODE WIRE3
:TRIGGER:ENHANCED:SPIBUS:MODE?
-> :TRIGGER:ENHANCED:SPIBUS:MODE WIRE3
:TRIGger:ENHanced:TV?
Function Queries all settings related to the TV trigger.
Syntax
:TRIGger:ENHanced:TV?
Example
:TRIGGER:ENHANCED:TV?
-> :TRIGGER:ENHANCED:TV:CUSTOMIZE 1;
FIELD DONTCARE;FRAME 2;HDTV:LINE 2;
POLARITY NEGATIVE;:TRIGGER:ENHANCED:
TV:LEVEL 1.000E+00;NTSC:LINE 5;
POLARITY NEGATIVE;:TRIGGER:ENHANCED:TV:
PAL:LINE 2;POLARITY NEGATIVE;:TRIGGER:
ENHANCED:TV:SDTV:LINE 8;
POLARITY NEGATIVE;:TRIGGER:ENHANCED:TV:
SGUARD 60;SOURCE 1;TYPE HDTV;
USERDEFINE:DEFINITION HD;
HFREJECTION OFF;HSYNC 50.00E+06;LINE 2;
POLARITY NEGATIVE
:TRIGger:ENHanced:TV:COUPling?
Function Queries the trigger coupling of the TV trigger.
Syntax
:TRIGger:ENHanced:TV:COUPling?
Example
:TRIGGER:ENHANCED:TV:COUPLING?
-> :TRIGGER:ENHANCED:TV:COUPLING TV
:TRIGger:ENHanced:TV:CUSTomize
Function Turns ON/OFF the sync guard function of the TV
trigger or queries the current setting.
Syntax
:TRIGger:ENHanced:TV:
CUSTomize {<Boolean>}
:TRIGger:ENHanced:TV:CUSTomize?
Example
:TRIGGER:ENHANCED:TV:CUSTOMIZE ON
:TRIGGER:ENHANCED:TV:CUSTOMIZE?
-> :TRIGGER:ENHANCED:TV:CUSTOMIZE 1
:TRIGger:ENHanced:TV:FIELd
Function Sets the field of the TV trigger or queries the current
setting.
Syntax
:TRIGger:ENHanced:TV:FIELd {DONTcare|
<NRf>}
:TRIGger:ENHanced:TV:FIELd?
<NRf> = 1 or 2
Example
:TRIGGER:ENHANCED:TV:FIELD DONTCARE
:TRIGGER:ENHANCED:TV:FIELD?
-> :TRIGGER:ENHANCED:TV:FIELD DONTCARE
:TRIGger:ENHanced:TV:FRAMe
Function Sets the frame skip function of the TV trigger or
queries the current setting.
Syntax
:TRIGger:ENHanced:TV:FRAMe {<NRf>}
:TRIGger:ENHanced:TV:FRAMe?
<NRf> = 1, 2, 4, or 8
Example
:TRIGGER:ENHANCED:TV:FRAME 2
:TRIGGER:ENHANCED:TV:FRAME?
-> :TRIGGER:ENHANCED:TV:FRAME 2
5.29 TRIGger Group