Loading...
5-192
IM 701310-17E
:TRIGger:ENHanced:SPIBus:CLOCk:SOURce
Function Sets the clock trace of the SPI bus trigger or queries
the current setting.
Syntax
:TRIGger:ENHanced:SPIBus:CLOCk:
SOURce {<NRf>}
:TRIGger:ENHanced:SPIBus:CLOCk:SOURce?
<NRf> = 1 to 4
Example
:TRIGGER:ENHANCED:SPIBUS:CLOCK:SOURCE 1
:TRIGGER:ENHANCED:SPIBUS:CLOCK:SOURCE?
-> :TRIGGER:ENHANCED:SPIBUS:CLOCK:
SOURCE 1
:TRIGger:ENHanced:SPIBus:CS?
Function Queries all settings related to the chip select of the
SPI bus trigger.
Syntax
:TRIGger:ENHanced:SPIBus:CS?
Example
:TRIGGER:ENHANCED:SPIBUS:CS?
-> :TRIGGER:ENHANCED:SPIBUS:CS:
ACTIVE HIGH;SOURCE 1
:TRIGger:ENHanced:SPIBus:CS:ACTive
Function Sets the active level of the chip select of the SPI bus
trigger or queries the current setting.
Syntax
:TRIGger:ENHanced:SPIBus:CS:
ACTive {HIGH|LOW}
:TRIGger:ENHanced:SPIBus:CS:ACTive?
Example
:TRIGGER:ENHANCED:SPIBUS:CS:ACTIVE HIGH
:TRIGGER:ENHANCED:SPIBUS:CS:ACTIVE?
-> :TRIGGER:ENHANCED:SPIBUS:CS:
ACTIVE HIGH
:TRIGger:ENHanced:SPIBus:CS:SOURce
Function Sets the chip select trace of the SPI bus trigger or
queries the current setting.
Syntax
:TRIGger:ENHanced:SPIBus:CS:
SOURce {<NRf>}
:TRIGger:ENHanced:SPIBus:CS:SOURce?
<NRf> = 1 to 4
Example
:TRIGGER:ENHANCED:SPIBUS:CS:SOURCE 1
:TRIGGER:ENHANCED:SPIBUS:CS:SOURCE?
-> :TRIGGER:ENHANCED:SPIBUS:CS:SOURCE 1
:TRIGger:ENHanced:SPIBus:DATA<x>?
Function Queries all settings related to the data of the SPI bus
trigger.
Syntax
:TRIGger:ENHanced:SPIBus:DATA<x>?
<x> = 1 or 2
Example
:TRIGGER:ENHANCED:SPIBUS:DATA1? -> :
TRIGGER:ENHANCED:SPIBUS:DATA1:BYTE 1;
CONDITION TRUE;DPOSITION 1;
PATTERN1 "00010010";
PATTERN2 "00110100";
PATTERN3 "01010110";
PATTERN4 "00010010";SOURCE 3
Description DAT
A2 is valid when
:TRIGger:ENHanced:
SPIBus:MODE WIRE4
is specified.
:TRIGger:ENHanced:SPIBus:DATA<x>:BYTE
Function Sets the number of bytes of the data of the SPI bus
trigger or queries the current setting.
Syntax
:TRIGger:ENHanced:SPIBus:DATA<x>:
BYTE {<NRf>}
:TRIGger:ENHanced:SPIBus:DATA<x>:BYTE?
<x> = 1 or 2
<NRf> = 1 to 4
Example
:TRIGGER:ENHANCED:SPIBUS:DATA1:BYTE 1
:TRIGGER:ENHANCED:SPIBUS:DATA1:BYTE?
-> :TRIGGER:ENHANCED:SPIBUS:DATA1:
BYTE 1
:TRIGger:ENHanced:SPIBus:DATA<x>:
CONDition
Function Sets the determination method (match or not match)
of the data of the SPI bus trigger or queries the
current setting.
Syntax
:TRIGger:ENHanced:SPIBus:DATA<x>:
CONDition {FALSe|TRUE}
:TRIGger:ENHanced:SPIBus:DATA<x>:
CONDition?
<x> = 1 or 2
Example
:TRIGGER:ENHANCED:SPIBUS:DATA1:
CONDITION TRUE
:TRIGGER:ENHANCED:SPIBUS:DATA1:
CONDITION? -> :TRIGGER:ENHANCED:SPIBUS:
DATA1:CONDITION TRUE
:TRIGger:ENHanced:SPIBus:DATA<x>:
DPOSition
Function Sets the pattern comparison start position of the data
of the SPI bus trigger or queries the current setting.
Syntax
:TRIGger:ENHanced:SPIBus:DATA<x>:
DPOSition {<NRf>}
:TRIGger:ENHanced:SPIBus:DATA<x>:
DPOSition?
<x> = 1 or 2
<NRf> = 0 to 9999
Example
:TRIGGER:ENHANCED:SPIBUS:DATA1:
DPOSITION 1
:TRIGGER:ENHANCED:SPIBUS:DATA1:
DPOSITION? -> :TRIGGER:ENHANCED:SPIBUS:
DATA1:DPOSITION 1
5.29 TRIGger Group