5-72
IM 710105-17E
:ANALysis:PMEAsure<x>:SETup:I:PROBe
Function Queries all current-to-voltage conversion ratio
settings of the probe connected to the current
input channel for power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:I:PRO
Be {C0_001|C0_002|C0_005|C0_01|C0_02|
C0_05|C0_1|C0_2|C0_5|C1|C2|C5|C10|
C20|C50|C100|C200|C500|C1000|C2000}
:ANALysis:PMEAsure<x>:SETup:I:PROBe?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1:SETUP:I:PROBE C0_
001
:ANALYSIS:PMEASURE1:SETUP:I:PROBE?
-> :ANALYSIS:PMEASURE1:SETUP:I:PROBE
C0_001
:ANALysis:PMEAsure<x>:SETup:RTRace
Function Sets or queries the deskewing source trace for
power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:RTRa
ce {I|U}
:ANALysis:PMEAsure<x>:SETup:RTRace?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1:SETUP:RTRACE I
:ANALYSIS:PMEASURE1:SETUP:RTRACE?
-> :ANALYSIS:PMEASURE1:SETUP:RTRACE I
:ANALysis:PMEAsure<x>:SETup:U?
Function Queries all voltage input channel settings for
power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:U?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1:SETUP:U?
:ANALysis:PMEAsure<x>:SETup:U:DESKew
Function Set or queries auto deskewing of the voltage input
channel for power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:U:DESK
ew {<Time>}
:ANALysis:PMEAsure<x>:SETup:U:DESKew?
<x> = 1 to 2
<Time>=−100.0nsto100.0ns(in10-pssteps)
Example
:ANALYSIS:PMEASURE1:SETUP:U:DESK
EW 1NS
:ANALYSIS:PMEASURE1:SETUP:U:DESKEW?
-> :ANALYSIS:PMEASURE1:SETUP:U:DESK
EW1.000E−09
:ANALysis:PMEAsure<x>:SETup:U:PROBe
Function Queries all voltage-to-voltage conversion ratio
settings of the probe connected to the voltage
input channel for power measurement.
Syntax
:ANALysis:PMEAsure<x>:SETup:U:PRO
Be {<NRf>}
:ANALysis:PMEAsure<x>:SETup:U:PROBe?
<x> = 1 to 2
<NRf> = See the DLM2000 Features Guide for
this information.
Example
:ANALYSIS:PMEASURE1:SETUP:U:PROBE 1
:ANALYSIS:PMEASURE1:SETUP:U:PROBE?
−>:ANALYSIS:PMEASURE1:SETUP:U:
PROBE 1.000
:ANALysis:PMEAsure<x>:TRANge
Function Sets or queries the measurement time period.
Syntax
:ANALysis:PMEAsure<x>:TRANge {<NRf>,<
NRf>}
:ANALysis:PMEAsure<x>:TRANge?
<x> = 1 to 2
<NRf>,<NRf>=−5to5divisions(instepsof10
divisions/display record length)
Example
:ANALYSIS:PMEASURE1:TRANGE5,−5
:ANALYSIS:PMEASURE1:TRANGE?
-> :ANALYSIS:PMEASURE1:TRANGE 5.00000
00E+00,−5.0000000E+00
:ANALysis:PMEAsure<x>:UDPRoximal?
Function Queries all distal, mesial, and proximal settings.
Syntax
:ANALysis:PMEAsure<x>:UDPRoximal?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1:UDPROXIMAL?
:ANALysis:PMEAsure<x>:UDPRoximal:MO
DE
Function Sets or queries the distal, mesial, and proximal
point mode setting.
Syntax
:ANALysis:PMEAsure<x>:UDPRoximal:MO
DE {PERCent|UNIT}
:ANALysis:PMEAsure<x>:UDPRoximal:MO
DE?
<x> = 1 to 2
Example
:ANALYSIS:PMEASURE1:UDPROXIMAL:MO
DE PERCENT
:ANALYSIS:PMEASURE1:UDPROXIMAL:MODE?
-> :ANALYSIS:PMEASURE1:UDPROXIMAL:MO
DE PERCENT
5.3 ANALysis Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals