EasyManuals Logo

YOKOGAWA DLM2000 Series User Manual

YOKOGAWA DLM2000 Series
340 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #129 background imageLoading...
Page #129 background image
5-96
IM 710105-17E
:FFT<x>:MEASure:PEAK[:BASic]:ALL
Function Collectively turns on or off all Basic peak items for
FFT analysis.
Syntax
:FFT<x>:MEASure:PEAK[:BASic]:A
LL {<Boolean>}
<x> = 1 or 2 (1 only for 2-channel models)
Example
:FFT1:MEASURE:PEAK:BASIC:ALL ON
:FFT<x>:MEASure:PEAK[:BASic]:DFReque
ncy?
Function Queries all of the settings for frequency
measurement between peak values for FFT
analysis.
Syntax
:FFT<x>:MEASure:PEAK[:BASic]:DFRequen
cy?
<x> = 1 or 2 (1 only for 2-channel models)
:FFT<x>:MEASure:PEAK[:BASic]:DFReque
ncy:STATe
Function Sets or queries the on/off status of the frequency
measured between peak values for FFT analysis.
Syntax
:FFT<x>:MEASure:PEAK[:BASic]:DFRequen
cy:STATe {<Boolean>}
:FFT<x>:MEASure:PEAK[:BASic]:DFRequen
cy:STATe?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:FFT1:MEASURE:PEAK:BASIC:DFREQUENCY:S
TATE ON
:FFT1:MEASURE:PEAK:BASIC:DFREQUENCY:S
TATE?
-> :FFT1:MEASURE:PEAK:BASIC:DFREQUENC
Y:STATE 1
:FFT<x>:MEASure:PEAK[:BASic]:DFReque
ncy:VALue?
Function Queries the frequency between peak values for
FFT analysis.
Syntax
:FFT<x>:MEASure:PEAK[:BASic]:DFRequen
cy:VALue?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:FFT1:MEASURE:PEAK:BASIC:DFREQUENCY:V
ALUE?
-> :FFT1:MEASURE:PEAK:BASIC:DFREQUENC
Y:VALUE 1.000E+00
:FFT<x>:MEASure:PEAK[:BASic]:DV?
Function Queries all of the settings for power measurement
between peak values for FFT analysis.
Syntax
:FFT<x>:MEASure:PEAK[:BASic]:DV?
<x> = 1 or 2 (1 only for 2-channel models)
:FFT<x>:MEASure:PEAK[:BASic]:DV:STA
Te
Function Sets or queries the on/off status of the power
measured between peak values for FFT analysis.
Syntax
:FFT<x>:MEASure:PEAK[:BASic]:DV:STA
Te {<Boolean>}
:FFT<x>:MEASure:PEAK[:BASic]:DV:STA
Te?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:FFT1:MEASURE:PEAK:BASIC:DV:STATE ON
:FFT1:MEASURE:PEAK:BASIC:DV:STATE?
-> :FFT1:MEASURE:PEAK:BASIC:DV:
STATE 1
:FFT<x>:MEASure:PEAK[:BASic]:DV:VAL
ue?
Function Queries the power between peak values for FFT
analysis.
Syntax
:FFT<x>:MEASure:PEAK[:BASic]:DV:VAL
ue?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:FFT1:MEASURE:PEAK:BASIC:DV:VALUE?
-> :FFT1:MEASURE:PEAK:BASIC:DV:
VALUE 1.000E+00
:FFT<x>:MEASure:PEAK[:BASic]:FREQuen
cy<y>?
Function Queries all of the settings for a peak frequency for
FFT analysis.
Syntax
:FFT<x>:MEASure:PEAK[:BASic]:FREQuenc
y<y>?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
:FFT<x>:MEASure:PEAK[:BASic]:FREQuen
cy<y>:STATe
Function Sets or queries the on/off status of a peak
frequency value for FFT analysis.
Syntax
:FFT<x>:MEASure:PEAK[:BASic]:FREQuenc
y<y>:STATe {<Boolean>}
:FFT<x>:MEASure:PEAK[:BASic]:FREQuenc
y<y>:STATe?
<x> = 1 or 2 (1 only for 2-channel models)
<y> = 1 or 2
Example
:FFT1:MEASURE:PEAK:BASIC:FREQUENCY1:S
TATE ON
:FFT1:MEASURE:PEAK:BASIC:FREQUENCY1:S
TATE?
-> :FFT1:MEASURE:PEAK:BASIC:FREQUENCY
1:STATE 1
5.11 FFT Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals