EasyManuals Logo

YOKOGAWA DLM2000 Series User Manual

YOKOGAWA DLM2000 Series
340 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
5-98
IM 710105-17E
:FFT<x>:TYPE
Function Sets or queries the spectrum for FFT analysis.
Syntax
:FFT<x>:TYPE {CH_MAG|CS_IMAG|
CS_LOGMAG|CS_MAG|CS_PHASE|CS_REAL|
LS_IMAG|LS_LOGMAG|LS_MAG|LS_PHASE|
LS_REAL|PS_LOGMAG|PS_MAG|PSD_LOGMAG|
PSD_MAG|RS_LOGMAG|RS_MAG|TF_IMAG|
TF_LOGMAG|TF_MAG|TF_PHASE|
TF_REAL[,<NRf>|MATH<x>]}
:FFT<x>:TYPE?
<x> of FFT<x> = 1 or 2 (1 only for 2-channel
models)
<NRf> = 1 to 4 (1 or 2 for 2-channel models)
<x> MATH<x> = 1 or 2 (1 only for 2-channel
models)
Example
:FFT1:TYPE CH_MAG,1
:FFT1:TYPE?
-> :FFT1:TYPE CH_MAG,1
Description This command results in an error on models
that are not equipped with the user-defined
computation option.
When the spectrum is set to (CH_MAG|CS_
IMAG|CS_LOGMAG|CS_MAG|CS_
PHASE|CS_REAL|TF_IMAG|TF_LOGMAG|TF_
MAG|TF_PHASE|TF_REAL), use <NRf> to
select the second waveform under analysis.
:FFT<x>:UNIT?
Function Queries all FFT computation unit settings.
Syntax
:FFT<x>:UNIT?
<x>= 1 or 2(1 only for the DLM2022, DLM2032,
DLM2052)
:FFT<x>:UNIT[:DEFine]
Function Sets or queries an FFT computation unit.
Syntax
:FFT<x>:UNIT[:DEFine] {<String>}
:FFT<x>:UNIT[:DEFine]?
<x>= 1 or 2(1 only for the DLM2022, DLM2032,
DLM2052)
Example
:FFT1:UNIT:DEFINE "EU"
:FFT1:UNIT:DEFINE?
-> :FFT1:UNIT:DEFINE "EU"
Description Units are applied to scale values. They never
affect the computation result.
:FFT<x>:UNIT:MODE
Function Sets or queries whether an FFT computation unit
will be attached automatically or manually.
Syntax
:FFT<x>:UNIT:MODE {AUTO|USERdefine}
:FFT<x>:UNIT:MODE?
<x>= 1 or 2(1 only for the DLM2022, DLM2032,
DLM2052)
Description
:FFT1:UNIT:MODE AUTO
:FFT1:UNIT:MODE?
-> :FFT1:UNIT:MODE AUTO
:FFT<x>:VERTical?
Function Queries all vertical axis settings for FFT analysis.
Syntax
:FFT<x>:VERTical?
<x> = 1 or 2 (1 only for 2-channel models)
:FFT<x>:VERTical:LEVel
Function Sets or queries the vertical axis display position
for FFT analysis.
Syntax
:FFT<x>:VERTical:LEVel {<NRf>}
:FFT<x>:VERTical:LEVel?
<x> = 1 or 2 (1 only for 2-channel models)
<NRf>=−1.0000E+31to1.0000E+31
Example
:FFT1:VERTICAL:LEVEL 1
:FFT1:VERTICAL:LEVEL?
-> :FFT1:VERTICAL:LEVEL 1.000E+00
:FFT<x>:VERTical:MODE
Function Sets or queries the FFT analysis vertical axis
mode.
Syntax
:FFT<x>:VERTical:MODE {AUTO|MANual}
:FFT<x>:VERTical:MODE?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:FFT1:VERTICAL:MODE AUTO
:FFT1:VERTICAL:MODE?
-> :FFT1:VERTICAL:MODE AUTO
:FFT<x>:VERTical:SENSitivity
Function Sets or queries the vertical sensitivity for FFT
analysis.
Syntax
:FFT<x>:VERTical:SENSitivity {<NRf>}
:FFT<x>:VERTical:SENSitivity?
<x> = 1 or 2 (1 only for 2-channel models)
<NRf>=−1.0000E+31to1.0000E+31
Example
:FFT1:VERTICAL:SENSITIVITY 1
:FFT1:VERTICAL:SENSITIVITY?
-> :FFT1:VERTICAL:SENSITIVITY 1.000E+
00
:FFT<x>:VTDisplay
Function Sets or queries the on/off status of the VT
waveform display for FFT analysis.
Syntax
:FFT<x>:VTDisplay {<Boolean>}
:FFT<x>:VTDisplay?
<x> = 1 or 2 (1 only for 2-channel models)
Example
:FFT1:VTDISPLAY ON
:FFT1:VTDISPLAY?
-> :FFT1:VTDISPLAY 1
5.11 FFT Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals