5-142
IM 710105-17E
:MEASure:HISTory:ABORt
Function Aborts statistical processing of automatically
measured history waveforms.
Syntax
:MEASure:HISTory:ABORt
Example
:MEASURE:HISTORY:ABORT
:MEASure:HISTory:EXECute
Function Executes statistical processing of automatically
measured history waveforms.
Syntax
:MEASure:HISTory:EXECute
Example
:MEASURE:HISTORY:EXECUTE
:MEASure:INDicator
Function Sets or queries the measurement location
indicator.
Syntax
:MEASure:INDicator {<NRf>|BIT<x>|MATH
<x>|OFF{,<Parameter>,2}}
:MEASure:INDicator?
<NRf> = 1 to 4 (1 or 2 for 2-channel models)
<x> of BIT<x> = 1 or 8
<x> of MATH<x> = 1 or 2 (1 only for 2-channel
models)
<Parameter> = {AMPLitude|
AVERage|AVGFre
q|AVGPeriod|BWIDth|DELay|DUTYcycle|FA
LL|FREQuency|HIGH|LOW|MAXimum|MINimu
m|NOVershoot|NWIDth|P|PABS|PERiod|PN
|POVershoot|PP|PTOPeak|PWIDth|RISE|RM
S|V1|V2|WH|WHABs|WHN|WHP|Z
}
Example
:MEASURE:INDICATOR 1,AVERAGE
:MEASURE:INDICATOR?
-> :MEASURE:INDICATOR 1,AVERAGE
:MEASure:MODE
Function Sets or queries the on/off/statistical processing
status of automated measurement.
Syntax
:MEASure:MODE {OFF|ON|CONTinuous|CYCL
e|HISTory}
:MEASure:MODE?
Example
:MEASURE:MODE OFF
:MEASURE:MODE?
-> :MEASURE:MODE OFF
:MEASure:RANGe<x>
Function Sets or queries the measurement source window.
Syntax
:MEASure:RANGe<x> {MAIN|Z1|Z2}
:MEASure:RANGe<x>?
<x> = 1 or 2
Example
:MEASURE:RANGE1 MAIN
:MEASURE:RANGE1?
-> :MEASURE:RANGE1 MAIN
Description
:Measure:RANGe1
sets or queries the
measurement source window for normal
measurement or the measurement source
window of Area1 for two-area measurement.
:Measure:RANGe2
sets or queries the
measurement source window of Area2 for two-
area measurement.
:MEASure:TRANge<x> (Time Range)
Function Sets or queries a time range.
Syntax
:MEASure:TRANge<x> {<NRf>,<NRf>}
:MEASure:TRANge<x>?
<x> = 1 or 2
<NRf>=−5to5divisions(instepsof10divisions/
display record length)
Example
:MEASURE:TRANGE1−4,4
:MEASURE:TRANGE1?
-> :MEASURE:TRANGE1−4.00E+00,4.00E+0
0
Description
:Measure:TRANge1
sets or queries the time
range for normal measurement or the time
range of Area1 for two-area measurement.
:Measure:TRANge2
sets or queries the time
range of Area2 for two-area measurement.
:MEASure:USER<x>?
Function Queries all automatic measurement settings for a
Calc item.
Syntax
:MEASure:USER<x>?
<x> = 1 to 4
:MEASure:USER<x>:COUNt?
Function Queries the statistical processing count of the
automatically measured value of a Calc item.
Syntax
:MEASure:USER<x>:COUNt?
<x> = 1 to 4
Example
:MEASURE:USER1:COUNT?
-> :MEASURE:USER1:COUNT 100
:MEASure:USER<x>:DEFine
Function Sets or queries the expression for the
automatically measured value of a Calc item.
Syntax
:MEASure:USER<x>:DEFine {<String>}
:MEASure:USER<x>:DEFine?
<x> = 1 to 4
<String> = Up to 128 characters
Example
:MEASURE:USER1:DEFINE "MAX(C1)"
:MEASURE:USER1:DEFINE?
-> :MEASURE:USER1:DEFINE "MAX(C1)"
:MEASure:USER<x>:{MAXimum|MEAN|MINim
um|SDEViation}?
Function Queries a statistical value that is calculated on the
automatically measured value of a Calc item.
Syntax
:MEASure:USER<x>:MAXimum?
<x> = 1 to 4
Example (Below are examples for the maximum value.)
:MEASURE:USER1:MAXIMUM?
-> :MEASURE:USER1:MAXIMUM 1.000E+00
Description If the statistical value is immeasurable, the
DLM2000 returns “NAN” (not a number).
5.20 MEASure Group