5-181
IM 710105-17E
Commands
1
2
3
4
5
6
App
Index
:SERialbus<x>:I2C:SEARch:SETup:MODE
Function Sets or queries the I
2
C bus signal search type.
Syntax
:SERialbus<x>:I2C:SEARch:SETup:MO
DE {ADRData|EVERystart|GENeralcall|HS
Mode|NONack|STARtbyte}
:SERialbus<x>:I2C:SEARch:SETup:MODE?
<x> = 1 to 4
Example
:SERIALBUS1:I2C:SEARCH:SETUP:
MODE ADRDATA
:SERIALBUS1:I2C:SEARCH:SETUP:MODE?
-> :SERIALBUS1:I2C:SEARCH:SETUP:
MODE ADRDATA
:SERialbus<x>:I2C:SEARch:SETup:NONa
ck?
Function Queries all NON ACK ignore mode settings for I
2
C
bus signal searching.
Syntax
:SERialbus<x>:I2C:SEARch:SETup:NONa
ck?
<x> = 1 to 4
:SERialbus<x>:I2C:SEARch:SETup:NONac
k:HSMode
Function Sets or queries whether or not NON ACK will be
ignored in high-speed mode for I
2
C bus signal
searching.
Syntax
:SERialbus<x>:I2C:SEARch:SETup:
NONack:HSMode {<Boolean>}
:SERialbus<x>:I2C:SEARch:SETup:
NONack:HSMode?
<x> = 1 to 4
Example
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:H
SMODE ON
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:H
SMODE?
-> :SERIALBUS1:I2C:SEARCH:SETUP:NONAC
K:HSMODE 1
:SERialbus<x>:I2C:SEARch:SETup:NONac
k:READaccess
Function Sets or queries whether or not NON ACK will be
ignored in read access mode for I
2
C bus signal
searching.
Syntax
:SERialbus<x>:I2C:SEARch:SETup:
NONack:READaccess {<Boolean>}
:SERialbus<x>:I2C:SEARch:SETup:
NONack:READaccess?
<x> = 1 to 4
Example
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:R
EADACCESS ON
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:R
EADACCESS?
-> :SERIALBUS1:I2C:SEARCH:SETUP:NONAC
K:READACCESS 1
:SERialbus<x>:I2C:SEARch:SETup:NONac
k:STARtbyte
Function Sets or queries whether or not NON ACK will be
ignored in start bytes for I
2
C bus signal searching.
Syntax
:SERialbus<x>:I2C:SEARch:SETup:
NONack:STARtbyte {<Boolean>}
:SERialbus<x>:I2C:SEARch:SETup:
NONack:STARtbyte?
<x> = 1 to 4
Example
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:S
TARTBYTE ON
:SERIALBUS1:I2C:SEARCH:SETUP:NONACK:S
TARTBYTE?
-> :SERIALBUS1:I2C:SEARCH:SETUP:NONAC
K:STARTBYTE 1
:SERialbus<x>:LIN?
Function Queries all LIN bus signal analysis and search
settings.
Syntax
:SERialbus<x>:LIN?
<x> = 1 to 4
:SERialbus<x>:LIN[:ANALyze]?
Function Queries all LIN bus signal analysis settings.
Syntax
:SERialbus<x>:LIN[:ANALyze]?
<x> = 1 to 4
:SERialbus<x>:LIN[:ANALyze]:SETup?
Function Queries all LIN bus signal analysis bus settings.
Syntax
:SERialbus<x>:LIN[:ANALyze]:SETup?
<x> = 1 to 4
:SERialbus<x>:LIN[:ANALyze]:SETup:BR
ATe
Function Sets or queries the LIN bus signal analysis bit rate
(data transfer rate).
Syntax
:SERialbus<x>:LIN[:ANALyze]:SETup:BRA
Te {<NRf>|USER,<NRf>}
:SERialbus<x>:LIN[:ANALyze]:SETup:BRA
Te?
<x> = 1 to 4
<NRf> = 1200, 2400, 4800, 9600, 19200
USER <NRf> = See the DLM2000 Features Guide
for this information.
Example
:SERIALBUS1:LIN:ANALYZE:SETUP:BRA
TE 2400
:SERIALBUS1:LIN:ANALYZE:SETUP:BRATE?
-> :SERIALBUS1:LIN:ANALYZE:SETUP:BRA
TE 2400
5.24 SERialbus Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals