5-182
IM 710105-17E
:SERialbus<x>:LIN[:ANALyze]:SETup:RE
Vision
Function Sets or queries the LIN bus signal analysis
revision number.
Syntax
:SERialbus<x>:LIN[:ANALyze]:SETup:REV
ision {BOTH|LIN1_3|LIN2_0}
:SERialbus<x>:LIN[:ANALyze]:SETup:REV
ision?
<x> = 1 to 4
Example
:SERIALBUS1:LIN:ANALYZE:SETUP:REVISI
ON LIN1_3
:SERIALBUS1:LIN:ANALYZE:SETUP:REVISI
ON?
-> :SERIALBUS1:LIN:ANALYZE:SETUP:REVI
SION LIN1_3
:SERialbus<x>:LIN[:ANALyze]:SETup:SO
URce
Function Sets or queries the LIN bus signal analysis source.
Syntax
:SERialbus<x>:LIN[:ANALyze]:SETup:SOU
Rce {<NRf>|MATH<x>}
:SERialbus<x>:LIN[:ANALyze]:SETup:SOU
Rce?
<x> = 1 to 4
<NRf> = 1 to 4
<x> of MATH<x> = 1 or 2
Example
:SERIALBUS1:LIN:ANALYZE:SETUP:SOUR
CE 1
:SERIALBUS1:LIN:ANALYZE:SETUP:SOURCE?
-> :SERIALBUS1:LIN:ANALYZE:SETUP:SOUR
CE 1
:SERialbus<x>:LIN[:ANALyze]:SETup:SP
Oint
Function Sets or queries the LIN bus signal sample point.
Syntax
:SERialbus<x>:LIN[:ANALyze]:SETup:SPO
int {<NRf>}
:SERialbus<x>:LIN[:ANALyze]:SETup:SPO
int?
<x> = 1 to 4
<NRf> = 18.8 to 90.6
Example
:SERIALBUS1:LIN:ANALYZE:SETUP:SPOI
NT 18.8
:SERIALBUS1:LIN:ANALYZE:SETUP:SPOINT?
-> :SERIALBUS1:LIN:ANALYZE:SETUP:SPOI
NT 18.8
:SERialbus<x>:LIN:DETail?
Function Queries all LIN bus signal analysis result list
settings.
Syntax
:SERialbus<x>:LIN:DETail?
<x> = 1 to 4
:SERialbus<x>:LIN:DETail:DISPlay
Function Sets or queries the display mode for the LIN bus
signal analysis result list.
Syntax
:SERialbus<x>:LIN:DETail:DISPl
ay {FULL|LOWer|UPPer}
:SERialbus<x>:LIN:DETail:DISPlay?
<x> = 1 to 4
Example
:SERIALBUS1:LIN:DETAIL:DISPLAY FULL
:SERIALBUS1:LIN:DETAIL:DISPLAY?
-> :SERIALBUS1:LIN:DETAIL:DISPLAY FUL
L
:SERialbus<x>:LIN:DETail:LIST:ITEM?
Function Queries all items that will be displayed in the LIN
bus signal analysis result list.
Syntax
:SERialbus<x>:LIN:DETail:LIST:ITEM?
<x> = 1 to 4
Example
:SERIALBUS1:LIN:DETAIL:LIST:ITEM?
-> :SERIALBUS1:LIN:DETAIL:LIST:
ITEM "No.,Time(ms),ID,ID-Field,Data,
Checksum,Information"
:SERialbus<x>:LIN:DETail:LIST:VALue?
Function Queries all of the data for the specified analysis
number in the LIN bus signal analysis result list.
Syntax
:SERialbus<x>:LIN:DETail:LIST:
VALue? {<NRf>|MAXimum|MINimum}
<x> = 1 to 4
<NRf> = See the DLM2000 Features Guide for
this information.
Example
:SERIALBUS1:LIN:DETAIL:LIST:VALUE? 0
-> :SERIALBUS1:LIN:DETAIL:LIST:
VALUE"0,−1,0016,30,F0,2D04,CE,,"
:SERialbus<x>:LIN:SEARch?
Function Queries all LIN bus signal search settings.
Syntax
:SERialbus<x>:LIN:SEARch?
<x> = 1 to 4
:SERialbus<x>:LIN:SEARch:ABORt
Function Aborts the LIN bus signal search.
Syntax
:SERialbus<x>:LIN:SEARch:ABORt
<x> = 1 to 4
Example
:SERIALBUS1:LIN:SEARCH:ABORT
5.24 SERialbus Group