5-189
IM 710105-17E
Commands
5
:SERialbus<x>:PSI5[:ANALyze]:SETup:S
LOT<y>
Function Sets or queries the slot’s start and end times for
PSI5 signal analysis.
Syntax
:SERialbus<x>:PSI5[:ANALyze]:SETup:SL
OT<y> {<NRf>[,<NRf>]}
:SERialbus<x>:PSI5[:ANALyze]:SETup:SL
OT<y>?
<x> = 1 to 4
<y> = 1 to 6 (slot number)
<NRf> = 0 to 0.02 s (in 0.1 us steps)
Example
:SERIALBUS1:PSI5:ANALYZE:SETUP:SL
OT6 870us,1088us
:SERIALBUS1:PSI5:ANALYZE:SETUP:SLOT1?
-> :SERIALBUS1:PSI5:ANALYZE:SETUP:SL
OT1870.000E-06,1.08800E−03
Description The first <NRF> is the slot's start time and the
second the end time.
If
:SERialbus<x>:PSI5[:ANALyze]:SETup:
NUMBerofslot
is a number from 1 to 6, when the
last slot number is specified with <y>, the second
<NRf> becomes valid.
If
:SERialbus<x>:PSI5[:ANALyze]:SETup:
NUMBerofslot
is
AUTO
, query is not possible.
:SERialbus<x>:PSI5[:ANALyze]:SETup:S
YNC
Function Sets or queries the sync signal for PSI5 signal
analysis.
Syntax
:SERialbus<x>:PSI5[:ANALyze]:SETup:SY
NC {<NRf>|MATH<y>|NONE}
:SERialbus<x>:PSI5[:ANALyze]:SETup:SY
NC?
<x> = 1 to 4
<y> = 1 or 2
<NRf> = 1 to 4
Example
:SERIALBUS1:PSI5:ANALYZE:SETUP:SYNC 1
:SERIALBUS1:PSI5:ANALYZE:SETUP:SYNC?
-> :SERIALBUS1:PSI5:ANALYZE:SETUP:SY
NC 1
:SERialbus<x>:PSI5:DETail?
Function Queries all PSI5 signal analysis list settings.
Syntax
:SERialbus<x>:PSI5:DETail?
<x> = 1 to 4
:SERialbus<x>:PSI5:DETail:DISPlay
Function Sets or queries the display mode for the PSI5
signal analysis list.
Syntax
:SERialbus<x>:PSI5:DETail:DISPlay {FU
LL|LOWer|UPPer}
:SERialbus<x>:PSI5:DETail:DISPlay?
<x> = 1 to 4
Example
:SERIALBUS1:PSI5:DETAIL:DISPLAY FULL
:SERIALBUS1:PSI5:DETAIL:DISPLAY?
-> :SERIALBUS1:PSI5:DETAIL:DISPLAY FU
LL
:SERialbus<x>:PSI5:DETail:LIST:ALL?
Function Queries all the data for all analysis numbers in the
PSI5 signal analysis result list.
Syntax
:SERialbus<x>:PSI5:DETail:LIST:ALL?
<x> = 1 to 4
Example :SERIALBUS1:PSI5:DETAIL:LIST:ALL?
-> #8 (number of bytes, 8 digits)(data sequence)
Description All the data for all analysis numbers are returned
in <block data> format.
The data of each analysis number is separated by
ASCII code 0AH.
:SERialbus<x>:PSI5:DETail:LIST:ITEM?
Function Queries all items that will be displayed in the PSI5
signal analysis result list.
Syntax
:SERialbus<x>:PSI5:DETail:LIST:ITEM?
<x> = 1 to 4
Example
:SERIALBUS1:PSI5:DETAIL:LIST:ITEM?
-> :SERIALBUS1:PSI5:DETAIL:LIST:IT
EM "No.Time(ms)Slot No.DataParity/
CRCInformation"
:SERialbus<x>:PSI5:DETail:LIST:VAL
ue?
Function Queries all the data for the specified analysis
number in the PSI5 signal analysis result list.
Syntax
:SERialbus<x>:PSI5:DETail:LIST:VAL
ue? {<NRf>|MAXimum|MINimum}
<x> = 1 to 4
<NRf>=−400000to400000
Example
:SERIALBUS1:PSI5:DETAIL:LIST:VALUE? 1
-> :SERIALBUS1:PSI5:DETAIL:LIST:VAL
UE "1, 0.062104,1,AB67,2,,,"
:SERialbus<x>:PSI5:SEARch?
Function Queries all PSI5 signal search settings.
Syntax
:SERialbus<x>:PSI5:SEARch?
<x> = 1 to 4
:SERialbus<x>:PSI5:SEARch:ABORt
Function Aborts the PSI5 signal search.
Syntax
:SERialbus<x>:PSI5:SEARch:ABORt
<x> = 1 to 4
Example
:SERIALBUS1:PSI5:SEARCH:ABORT
:SERialbus<x>:PSI5:SEARch:EXECute
Function Executes a PSI5 signal search.
Syntax
:SERialbus<x>:PSI5:SEARch:EXECute
<x> = 1 to 4
Example
:SERIALBUS1:PSI5:SEARCH:EXECUTE
5.24 SERialbus Group