5-190
IM 710105-17E
:SERialbus<x>:PSI5:SEARch:SELect
Function Sets which detected point to display in the PSI5
signal search zoom window and queries the zoom
position of the detected point.
Syntax
:SERialbus<x>:PSI5:SEARch:SELe
ct {<NRf>|MAXimum}
:SERialbus<x>:PSI5:SEARch:SELect?
<x> = 1 to 4
<NRf> = 0 to 49999
Example
:SERIALBUS1:PSI5:SEARCH:SELECT 0
:SERIALBUS1:PSI5:SEARCH:SELECT?
-> :SERIALBUS1:PSI5:SEARCH:SELECT 0
Description If there are no detected points, the DLM2000
returns “NAN.”
:SERialbus<x>:PSI5:SEARch:SETup?
Function Queries all PSI5 signal search condition settings.
Syntax
:SERialbus<x>:PSI5:SEARch:SETup?
<x> = 1 to 4
:SERialbus<x>:PSI5:SEARch:SETup:DA
TA?
Function Queries all PSI5 signal data search settings.
Syntax
:SERialbus<x>:PSI5:SEARch:SETup:DATA?
<x> = 1 to 4
:SERialbus<x>:PSI5:SEARch:SETup:DATA
:DATA?
Function Queries all data settings of the PSI5 signal data
search.
Syntax
:SERialbus<x>:PSI5:SEARch:SETup:
DATA:DATA?
<x> = 1 to 4
:SERialbus<x>:PSI5:SEARch:SETup:DATA
:DATA:CONDition
Function Sets or queries the comparison condition of the
PSI5 signal data search.
Syntax
:SERialbus<x>:PSI5:SEARch:SETup:
DATA:DATA:CONDition {BETWeen|EQUal|FA
LSe|GREater|LESS|NOTBetween|NOTEqul|T
RUE}
:SERialbus<x>:PSI5:SEARch:SETup:
DATA:DATA:CONDition?
<x> = 1 to 4
Example
:SERIALBUS1:PSI5:SEARCH:SETUP:
DATA:DATA:CONDITION BETWEEN
:SERIALBUS1:PSI5:SEARCH:SETUP:
DATA:DATA:CONDITION?
-> :SERIALBUS1:PSI5:SEARCH:SETUP:
DATA:DATA:CONDITION BETWEEN
:SERialbus<x>:PSI5:SEARch:SETup:DATA
:DATA:DECimal<y>
Function Sets or queries the data of the PSI5 signal data
search in decimal notation.
Syntax
:SERialbus<x>:PSI5:SEARch:SETup:
DATA:DATA:DECimal<y> {<NRf>}
:SERialbus<x>:PSI5:SEARch:SETup:
DATA:DATA:DECimal<y>?
<x> = 1 to 4
<y> = 1 or 2
When<NRf>=10bitdata,−512to511
16bitdata,−32768to32767
Example
:SERIALBUS1:PSI5:SEARCH:SETUP:DATA:DA
TA:DECIMAL1−10
:SERIALBUS1:PSI5:SEARCH:SETUP:DATA:DA
TA:DECIMAL1?
-> :SERIALBUS1:PSI5:SEARCH:SETUP:
DATA:DATA:DECIMAL1−10
Description This command is valid when
:SERialbus<x>:PS
I5:SEARch:SETup:DATA:DATA:CONDition
is
BETWeen
,
EQUal
,
GREater
,
LESS
,
NOTBetween
,
or
NOTEqul
.
:SERialbus<x>:PSI5:SEARch:SETup:DATA
:DATA:HEXa
Function Sets the data of the PSI5 signal data search in
hexadecimal notation.
Syntax
:SERialbus<x>:PSI5:SEARch:SETup:
DATA:DATA:HEXa {<String>}
<x> = 1 to 4
Example
:SERIALBUS1:PSI5:SEARCH:SETUP:DATA:DA
TA:HEXA "3A0F"
Description This command is valid when
:SERialbus<x>:P
SI5:SEARch:SETup:DATA:DATA:CONDition
is
TRUE
or
FALSe
.
:SERialbus<x>:PSI5:SEARch:SETup:DATA
:DATA:PATTern
Function Sets or queries the data of the PSI5 signal data
search in binary notation.
Syntax
:SERialbus<x>:PSI5:SEARch:SETup:
DATA:DATA:PATTern {<String>}
:SERialbus<x>:PSI5:SEARch:SETup:
DATA:DATA:PATTern?
<x> = 1 to 4
Example
:SERIALBUS1:PSI5:SEARCH:SETUP:
DATA:DATA:PATTERN "0011010100001111"
:SERIALBUS1:PSI5:SEARCH:SETUP:
DATA:DATA:PATTERN?
-> :SERIALBUS1:PSI5:SEARCH:SETUP:
DATA:DATA:PATTERN "0011010100001111"
Description This command is valid when
:SERialbus<x>:P
SI5:SEARch:SETup:DATA:DATA:CONDition
is
TRUE
or
FALSe
.
5.24 SERialbus Group