EasyManuals Logo

YOKOGAWA DLM2000 Series User Manual

YOKOGAWA DLM2000 Series
340 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #231 background imageLoading...
Page #231 background image
5-198
IM 710105-17E
:SERialbus<x>:SENT:DETail:LIST:VAL
ue?
Function Queries all the data for the specified analysis
number in the SENT signal analysis result list.
Syntax
:SERialbus<x>:SENT:DETail:LIST:VAL
ue? {<NRf>|MAXimum|MINimum}
<x> = 1 to 4
<NRf> = –100000 to 100000
Example
:SERIALBUS1:SENT:DETail:LIST:VALUE? 0
-> :SERIALBUS1:SENT:DETAIL:LIST:VAL
UE "0,-0.1672,168.00,3.00,1100,6,3,
5,E,B,9,3,283.73,,,"
Description The items displayed in the analysis result list
varies depending on the :SERialbus<x>:SENT[:
ANALyze]:SETup:DISPlay {BOTH|FAST|SLOW}
setting.
:SERialbus<x>:SENT:SEARch?
Function Queries all SENT signal search settings.
Syntax
:SERialbus<x>:SENT:SEARch?
<x> = 1 to 4
:SERialbus<x>:SENT:SEARch:ABORt
Function Aborts the SENT signal search.
Syntax
:SERialbus<x>:SENT:SEARch:ABORt
<x> = 1 to 4
Example
:SERIALBUS1:SENT:SEARCH:ABORT
:SERialbus<x>:SENT:SEARch:EXECute
Function Executes a SENT signal search.
Syntax
:SERialbus<x>:SENT:SEARch:EXECute
<x> = 1 to 4
Example
:SERIALBUS1:SENT:SEARCH:EXECUTE
:SERialbus<x>:SENT:SEARch:SELect
Function Sets which detected point to display in the SENT
signal search zoom window and queries the zoom
position of the detected point.
Syntax
:SERialbus<x>:SENT:SEARch:SELe
ct {<NRf>|MAXimum}
:SERialbus<x>:SENT:SEARch:SELect?
<x> = 1 to 4
<NRf> = 0 to 49999
Example
:SERIALBUS1:SENT:SEARCH:SELECT 1
:SERIALBUS1:SENT:SEARCH:SELECT?
-> :SERIALBUS1:SENT:SEARCH:SELECT 1.5
00E+00
Description If there are no detected points, the DLM2000
returns “NAN.”
:SERialbus<x>:SENT:SEARch:SELe
ct? MAXimum
Function Queries the number of detected points in the
SENT signal search.
Syntax
:SERialbus<x>:SENT:SEARch:SELe
ct? {MAXimum}
<x> = 1 to 4
Example
:SERIALBUS1:SENT:SEARCH:SELECT? MAXim
um
-> :SERIALBUS1:SENT:SEARCH:SELECT 100
Description If there are no detected points, the DLM2000
returns “NAN.”
:SERialbus<x>:SENT:SEARch[:SETup]?
Function Queries all SENT signal search condition settings.
Syntax
:SERialbus<x>:SENT:SEARch[:SETup]?
<x> = 1 to 4
:SERialbus<x>:SENT:SEARch[:SETup]:ER
Ror?
Function Queries all SENT signal search error settings.
Syntax
:SERialbus<x>:SENT:SEARch:[SETup]:ERR
or?
<x> = 1 to 4
:SERialbus<x>:SENT:SEARch:[SETup]:ER
Ror:FCRC
Function Sets or queries the SENT signal search fast
channel CRC error.
Syntax
:SERialbus<x>:SENT:SEARch:[SETup]:ERR
or:FCRC {<Boolean>}
:SERialbus<x>:SENT:SEARch:[SETup]:ERR
or:FCRC?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:SEARCH:SETUP:ERROR:F
CRC ON
:SERIALBUS1:SENT:SEARCH:SETUP:ERROR:F
CRC?
-> :SERIALBUS1:SENT:SEARCH:SETUP:ERRO
R:FCRC 1
5.24 SERialbus Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals