EasyManuals Logo

YOKOGAWA DLM2000 Series User Manual

YOKOGAWA DLM2000 Series
340 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #230 background imageLoading...
Page #230 background image
5-197
IM 710105-17E
Commands
1
2
3
4
5
6
App
Index
:SERialbus<x>:SENT[:ANALyze]:SETup:F
ORMat:VERSion
Function Sets or queries the SENT signal analysis
specification version.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:FO
RMat:VERSion {FEB2008|JAN2010}
:SERialbus<x>:SENT[:ANALyze]:SETup:FO
RMat:VERESion?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMAT
:VERSION JAN2010
:SERIALBUS1:SENT:ANALYZE:SETUP:FORMAT
:VERSION?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:FOR
MAT:VERSION JAN2010
:SERialbus<x>:SENT[:ANALyze]:SETup:S
TYPe
Function Sets or queries the SENT signal analysis slow
channel format.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:ST
YPe {ENHanced|SHORt}
:SERialbus<x>:SENT[:ANALyze]:SETup:ST
YPe?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:STY
PE ENHANCED
:SERIALBUS1:SENT:ANALYZE:SETUP:STYPE?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:STY
PE ENHANCED
:SERialbus<x>:SENT[:ANALyze]:SETup:S
OURce
Function Sets or queries the SENT signal analysis source.
Syntax
:SERialbus<x>:SENT[:ANALyze]:SETup:SO
URce {<NRf>|BIT<x>|MATH<x>}
:SERialbus<x>:SENT[:ANALyze]:SETup:SO
URce?
<x> = 1 to 4
<NRf> = 1 to 8
<x> of BIT<x> = 1 to 8
<x> of MATH<x> = 1 or 2
Example
:SERIALBUS1:SENT:ANALYZE:SETUP:SOUR
CE 1
:SERIALBUS1:SENT:ANALYZE:SETUP:SOUR
CE?
-> :SERIALBUS1:SENT:ANALYZE:SETUP:SOU
RCE 1
:SERialbus<x>:SENT:DETail?
Function Queries all SENT signal analysis list settings.
Syntax
:SERialbus<x>:SENT:DETail?
<x> = 1 to 4
:SERialbus<x>:SENT:DETail:DISPlay
Function Sets or queries the display mode for the SENT
signal analysis list.
Syntax
:SERialbus<x>:SENT:DETail:DISPlay {FU
LL|LOWer|UPPer}
:SERialbus<x>:SENT:DETail:DISPlay?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:DETAIL:DISPLAY FULL
:SERIALBUS1:SENT:DETAIL:DISPLAY?
-> :SERIALBUS1:SENT:DETAIL:DISPLAY FU
LL
:SERialbus<x>:SENT:DETail:LIST:ALL?
Function Queries all the data for all analysis numbers in the
SENT signal analysis result list.
Syntax
:SERialbus<x>:SENT:DETail:LIST:ALL?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:DETAIL:LIST:ALL?
-> #8
(number of bytes, 8 digits)(data sequence)
Description All the data for all analysis numbers are returned
in <block data> format.
The data of each analysis number is separated by
ASCII code 0AH.
:SERialbus<x>:SENT:DETail:LIST:ITEM?
Function Queries all items that will be displayed in the
SENT signal analysis result list.
Syntax
:SERialbus<x>:SENT:DETail:LIST:ITEM?
<x> = 1 to 4
Example
:SERIALBUS1:SENT:DETail:LIST:ITEM?
-> :SERIALBUS1:SENT:DETAIL:LIST:
ITEM "No.,Time(ms),Sync(us),Tick(us),
S&C,Data,CRC,Length(tick),Information
,SlowCH,"
Description The items displayed in the analysis result list
varies depending on the :SERialbus<x>:SENT[:
ANALyze]:SETup:DISPlay {BOTH|FAST|SLOW}
setting.
5.24 SERialbus Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals