5-258
IM 710105-17E
:TRIGger{[:ATRigger]|:BTRigger}:SIMP
le:COUPling
Function Sets or queries the edge trigger source trigger
coupling.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:COUPling {AC|DC}
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:COUPling?
Example
:TRIGGER:ATRIGGER:SIMPLE:COUPLING AC
:TRIGGER:ATRIGGER:SIMPLE:COUPLING?
-> :TRIGGER:ATRIGGER:SIMPLE:
COUPLING AC
Description You cannot use this command when the trigger
source is set to EXTernal, LINE, or LOGIC.
:TRIGger{[:ATRigger]|:BTR
igger}:SIMPle:HFRejection
(HighFrequencyREJECTION)
Function Sets or queries the on/off status of the trigger
source low-pass filter (HF rejection) for edge
triggering.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:HFRejection {<Frequency>|OFF}
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:HFRejection?
<Frequency> = 15 kHz to 20MHz
Example
:TRIGGER:ATRIGGER:SIMPLE:
HFREJECTION 15KHZ
:TRIGGER:ATRIGGER:SIMPLE:HFREJECTION?
-> :TRIGGER:ATRIGGER:SIMPLE:HFREJECTI
ON 15000
Description You cannot use this command when the trigger
source is set to EXTernal, LINE, or LOGIC.
:TRIGger{[:ATRigger]|:BTRigger}:SIMP
le:HYSTeresis
Function Sets or queries the noise rejection setting for the
edge trigger source trigger level.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:HYSTeresis {HIGH|LOW}
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:HYSTeresis?
Example
:TRIGGER:ATRIGGER:SIMPLE:HYSTERES
IS HIGH
:TRIGGER:ATRIGGER:SIMPLE:HYSTERESIS?
-> :TRIGGER:ATRIGGER:SIMPLE:HYSTERES
IS HIGH
Description You cannot use this command when the trigger
source is set to EXTernal, LINE, or LOGIC.
:TRIGger{[:ATRigger]|:BTRigger}:SIMP
le:LEVel
Function Sets or queries the edge trigger source trigger
level.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:LEVel {<Voltage>}
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:LEVel?
<Voltage> = See the DLM2000 Features Guide for
this information.
Example
:TRIGGER:ATRIGGER:SIMPLE:LEVEL 0V
:TRIGGER:ATRIGGER:SIMPLE:LEVEL?
-> :TRIGGER:ATRIGGER:SIMPLE:
LEVEL 0.000E+00
Description You cannot use this command when the trigger
source is set to LINE, or LOGIC.
When the probe type is set to current, this
command sets or queries the <Current> value.
:TRIGger{[:ATRigger]|:BTRigger}:SIMP
le:PROBe
Function Sets or queries the external trigger source probe
setting for edge triggering.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:PROBe {<NRf>}
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:PROBe?
<NRf> = 1 to 10
Example
:TRIGGER:ATRIGGER:SIMPLE:PROBE 1
:TRIGGER:ATRIGGER:SIMPLE:PROBE?
-> :TRIGGER:ATRIGGER:SIMPLE:PROBE 1
Description You can only use this command when the trigger
source is set to EXTernal.
:TRIGger{[:ATRigger]|:BTRigger}:SIMP
le:RANGe
Function Sets or queries the external trigger source input
range for edge triggering.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:RANGe {<Voltage>}
:TRIGger{[:ATRigger]|:BTRigger}:SIMPl
e:RANGe?
<Voltage> = 1 to 100 (V)
Example
:TRIGGER:ATRIGGER:SIMPLE:RANGE 1V
:TRIGGER:ATRIGGER:SIMPLE:RANGE?
-> :TRIGGER:ATRIGGER:SIMPLE:RANGE 1.0
E+00
Description You can only use this command when the
trigger source is set to EXTernal.
This command can be used on the DLM2022,
DLM2032, and DLM2052.
5.33 TRIGger Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals