Page #296 background imageLoading...
Page #296 background image
5-263
IM 710105-17E
Commands
1
2
3
4
5
6
App
Index
:TRIGger{[:ATRigger]|:BTRigger}:SPI:
DATA<x>:BCOunt
Function Sets or queries the starting position of data
comparison for SPI bus signal triggering.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:BCOunt {<NRf>}
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:BCOunt?
<x> = 1 or 2
<NRf> = 0 to 9999
Example
:TRIGGER:ATRIGGER:SPI:DATA1:BCOUNT 0
:TRIGGER:ATRIGGER:SPI:DATA1:BCOUNT?
-> :TRIGGER:ATRIGGER:SPI:DATA1:BCOU
NT 0
:TRIGger{[:ATRigger]|:BTRigger}:SPI:
DATA<x>:CONDition
Function Sets or queries the data comparison condition (true
or false) for SPI bus signal triggering.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:CONDition {FALSe|TRUE}
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:CONDition?
<x> = 1 or 2
Example
:TRIGGER:ATRIGGER:SPI:DATA1:CONDITI
ON FALSE
:TRIGGER:ATRIGGER:SPI:DATA1:CONDITI
ON?
-> :TRIGGER:ATRIGGER:SPI:DATA1:CONDIT
ION FALSE
:TRIGger{[:ATRigger]|:BTRigger}:SPI:
DATA<x>:DBYTe
Function Sets or queries the data size (in bytes) for SPI bus
signal triggering.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:DBYTe {<NRf>}
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:DBYTe?
<x> = 1 or 2
<NRf> = 1 to 4
Example
:TRIGGER:ATRIGGER:SPI:DATA1:DBYTE 1
:TRIGGER:ATRIGGER:SPI:DATA1:DBYTE?
-> :TRIGGER:ATRIGGER:SPI:DATA1:
DBYTE 1
:TRIGger{[:ATRigger]|:BTRigger}:SPI:
DATA<x>:HEXa<y>
Function Sets a data value for SPI bus signal triggering in
hexadecimal notation.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:HEXa<y> {<String>}
<x> = 1 or 2
<y> = 1 to 4
Example
:TRIGGER:ATRIGGER:SPI:DATA1:
HEXA1 "AB"
:TRIGger{[:ATRigger]|:BTRigger}:SPI:
DATA<x>:MODE
Function Sets or queries the data enable/disable condition
for SPI bus signal triggering (on/off).
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:MODE {<Boolean>}
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:MODE?
<x> = 1 or 2
Example
:TRIGGER:ATRIGGER:SPI:DATA1:MODE ON
:TRIGGER:ATRIGGER:SPI:DATA1:MODE?
-> :TRIGGER:ATRIGGER:SPI:DATA1:MODE 1
:TRIGger{[:ATRigger]|:BTRigger}:SPI:
DATA<x>:PATTern<y>
Function Sets or queries the data value for SPI bus signal
triggering in binary notation.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:PATTern<y> {<String>}
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:PATTern<y>?
<x> = 1 or 2
<y> = 1 to 4
Example
:TRIGGER:ATRIGGER:SPI:DATA1:PATTE
RN1 "11001010"
:TRIGGER:ATRIGGER:SPI:DATA1:PATTERN1?
-> :TRIGGER:ATRIGGER:SPI:DATA1:PATTE
RN1 "11001010"
:TRIGger{[:ATRigger]|:BTRigger}:SPI:
DATA<x>:PFORmat
Function Sets or queries the data input format, which is
one of the data conditions, for SPI bus signal
triggering.
Syntax
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:PFORmat {BINary|HEXa}
:TRIGger{[:ATRigger]|:BTRigger}:SPI:D
ATA<x>:PFORmat?
<x> = 1 or 2
Example
:TRIGGER:ATRIGGER:SPI:DATA1:PFORM
AT BINARY
:TRIGGER:ATRIGGER:SPI:DATA1:PFORMAT?
-> :TRIGGER:ATRIGGER:SPI:DATA1:PFORM
AT BINARY
5.33 TRIGger Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals