5-62
IM 710105-17E
:ANALysis:PANalyze<x>:HARMonics:SPOi
nt
Function Sets or queries the harmonic analysis computation
start point.
Syntax
:ANALysis:PANalyze<x>:HARMonics:SPOi
nt {<NRf>}
:ANALysis:PANalyze<x>:HARMonics:SPOi
nt?
<x> = 1 or 2
<NRf>=−5to5(divisions)
Example
:ANALYSIS:PANALYZE1:HARMONICS:SPOI
NT 1
:ANALYSIS:PANALYZE1:HARMONICS:SPOINT?
-> :ANALYSIS:PANALYZE1:HARMONICS:SPOI
NT 1.000E+00
:ANALysis:PANalyze<x>:HARMonics:SVOL
tage
Function Sets or queries the power supply voltage for
harmonic analysis.
Syntax
:ANALysis:PANalyze<x>:HARMonics:SVOLt
age {<NRf>|<Current>}
:ANALysis:PANalyze<x>:HARMonics:SVOLt
age?
<x> = 1 or 2
<NRf>, <Voltage> = 90 to 440 V
Example
:ANALYSIS:PANALYZE1:HARMONICS:SVOLTA
GE 230
:ANALYSIS:PANALYZE1:HARMONICS:SVOLTA
GE?
-> :ANALYSIS:PANALYZE1:HARMONICS:SVOL
TAGE 220.00000E+00
:ANALysis:PANalyze<x>:SETup?
Function Queries all power supply analysis input settings.
Syntax
:ANALysis:PANalyze<x>:SETup?
<x> = 1 or 2
:ANALysis:PANalyze<x>:SETup:ADESkew
Function Executes automatic deskewing for power supply
analysis.
Syntax
:ANALysis:PANalyze<x>:SETup:ADESkew
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:SETUP:ADESKEW
:ANALysis:PANalyze<x>:HARMonics:DETa
il:LIST:ITEM?
Function Queries the items displayed in the harmonic
analysis result list.
Syntax
:ANALysis:PANalyze<x>:HARMonics:DETai
l:LIST:ITEM?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:HARMONICS:DETAIL:
LIST:ITEM?
-> :ANALYSIS:PANALYZE1:HARMONICS:DETA
IL:LIST:ITEM "Order.,Measure(A),
Limit(A),Measure(%),Limit(%),Info,"
:ANALysis:PANalyze<x>:HARMonics:DETa
il:LIST:VALue?
Function Queries all of the data for the specified analysis
number in the harmonic analysis result list.
Syntax
:ANALysis:PANalyze<x>:HARMonics:DETai
l:LIST:VALue? {<NRf>}
<x> = 1 or 2
<NRf> = 1 to 40 (harmonic order)
Example
:ANALYSIS:PANALYZE1:HARMONICS:DETAIL:
LIST:VALUE? 2
-> :ANALYSIS:PANALYZE1:HARMONICS:DETA
IL:LIST:VALUE " 2, 0.031, 0.020, 3.14
9, 2.000,NG,"
:ANALysis:PANalyze<x>:HARMonics:GROu
ping
Function Sets or queries the harmonic analysis grouping.
Syntax
:ANALysis:PANalyze<x>:HARMonics:GROup
ing {OFF|TYPE1|TYPE2}
:ANALysis:PANalyze<x>:HARMonics:GROup
ing?
<x> = 1 or 2
Example
:ANALYSIS:PANALYZE1:HARMONICS:GROUPI
NG OFF
:ANALYSIS:PANALYZE1:HARMONICS:GROUPI
NG?
-> :ANALYSIS:PANALYZE1:HARMONICS:GROU
PING OFF
5.3 ANALysis Group

Table of Contents

Other manuals for YOKOGAWA DLM2000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM2000 Series and is the answer not in the manual?

YOKOGAWA DLM2000 Series Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM2000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals