EasyManuals Logo

YOKOGAWA DLM4038 User Manual

YOKOGAWA DLM4038
373 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #115 background imageLoading...
Page #115 background image
5-82
IM DLM4038-17EN
:CHANnel<x>:LSCale:UNIT
Function Sets or queries a unit that is attached to linear
scaling results.
Syntax
:CHANnel<x>:LSCale:UNIT {<String>}
:CHANnel<x>:LSCale:UNIT?
<x> = 1 to 8
<String> = Up to four characters
Example
:CHANNEL1:LSCALE:UNIT "EU"
:CHANNEL1:LSCALE:UNIT?
-> :CHANNEL1:LSCALE:UNIT "EU"
Description You can only use the characters and symbols
that appear on the DLM4000 soft keyboard.
Assignment of a unit never affects the scale
value.
:CHANnel<x>:OFFSet
Function Sets or queries the offset voltage of a channel.
Syntax
:CHANnel<x>:OFFSet {<Voltage>|<Curre
nt>}
:CHANnel<x>:OFFSet?
<x> = 1 to 8
<Voltage>, <Current> = See the DLM4000
Features Guide for this information.
Example
:CHANNEL1:OFFSET 0V
:CHANNEL1:OFFSET?
-> :CHANNEL1:OFFSET 0.00E+00
:CHANnel<x>:POSition
Function Sets or queries the vertical position of a channel.
Syntax
:CHANnel<x>:POSition {<NRf>}
:CHANnel<x>:POSition?
<x> = 1 to 8
<NRf> = − to 4 (divisions)
Example
:CHANNEL1:POSITION 1
:CHANNEL1:POSITION?
-> :CHANNEL1:POSITION 1.00E+00
:CHANnel<x>:PROBe?
Function Queries all probe attenuation settings of a
channel.
Syntax
:CHANnel<x>:PROBe?
<x> = 1 to 8
:CHANnel<x>:PROBe:DZCalibrate
Function Performs demagnetization and zero adjustment
on each channel.
Syntax
:CHANnel<x>:PROBe:DZCalibrate
<x> = 1 to 8
Example
:CHANNEL1:PROBE:DZCALIBRATE
:CHANnel<x>:LABel:DISPlay
Function Sets or queries the waveform label display on/off
status of a channel.
Syntax
:CHANnel<x>:LABel:DISPlay {<Boolean>}
:CHANnel<x>:LABel:DISPlay?
<x> = 1 to 8
Example
:CHANNEL1:LABEL:DISPLAY ON
:CHANNEL1:LABEL:DISPLAY?
-> :CHANNEL1:LABEL:DISPLAY 1
:CHANnel<x>:LSCale?
Function Queries all linear scaling settings of a channel.
Syntax
:CHANnel<x>:LSCale?
<x> = 1 to 8
:CHANnel<x>:LSCale:AVALue
Function Sets or queries a scaling coefficient A value.
Syntax
:CHANnel<x>:LSCale:AVALue {<NRf>}
:CHANnel<x>:LSCale:AVALue?
<x> = 1 to 8
<NRf> = −1.0000E+31 to 1.0000E+31
Example
:CHANNEL1:LSCALE:AVALUE 10
:CHANNEL1:LSCALE:AVALUE?
-> :CHANNEL1:LSCALE:AVALUE 10.0000E+
00
:CHANnel<x>:LSCale:BVALue
Function Sets or queries an offset B value.
Syntax
:CHANnel<x>:LSCale:BVALue {<NRf>}
:CHANnel<x>:LSCale:BVALue?
<x> = 1 to 8
<NRf> = −1.0000E+31 to 1.0000E+31
Example
:CHANNEL1:LSCALE:BVALUE 10
:CHANNEL1:LSCALE:BVALUE?
-> :CHANNEL1:LSCALE:BVALUE 10.0000E+
00
:CHANnel<x>:LSCale:MODE
Function Sets or queries the on/off status of linear scaling.
Syntax
:CHANnel<x>:LSCale:MODE {<Boolean>}
:CHANnel<x>:LSCale:MODE?
<x> = 1 to 8
Example
:CHANNEL1:LSCALE:MODE ON
:CHANNEL1:LSCALE:MODE?
-> :CHANNEL1:LSCALE:MODE 1
5.6 CHANnel Group

Table of Contents

Other manuals for YOKOGAWA DLM4038

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM4038 and is the answer not in the manual?

YOKOGAWA DLM4038 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM4038
CategoryTest Equipment
LanguageEnglish

Related product manuals