Page #270 background imageLoading...
Page #270 background image
5-237
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
5.27 SSTart Group
:SSTart?
Function Sets the trigger mode to SINGLE, and starts
waveform acquisition. Returns zero if data
acquisition is stopped before the specified time
period. Returns 1 if waveform acquisition is not
stopped within the specified time period.
Syntax
:SSTart? {<NRf>}
<NRf> = 1 to 36000 (in 100-ms steps. Start and
wait.)
<NRf> = 0 (start without any waiting)
<NRf> = 36000 to −1 (in 100-ms steps. Wait
without starting.)
Example
:SSTART? 10
-> :SSTART 1
Description If you specify a positive time value, the
DLM4000 starts waveform acquisition in Single
Trigger mode and waits until the waveform
acquisition is stopped within the specified time
period.
If you set the time value to zero, the DLM4000
starts waveform acquisition and returns zero
without waiting until the waveform acquisition is
stopped.
If you specify a negative time value, the
DLM4000 does not start waveform acquisition
and simply waits until the waveform acquisition
in progress is stopped within the specified time
period.

Table of Contents

Other manuals for YOKOGAWA DLM4038

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM4038 and is the answer not in the manual?

YOKOGAWA DLM4038 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM4038
CategoryTest Equipment
LanguageEnglish

Related product manuals