Loading...
5-129
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
:LOGic:PODB:BIT<x>:DISPlay
Function Turns on or off all bit displays of logic input port B.
Syntax
:LOGic:PODB:BIT<x>:DISPlay {<Boolean
>}
:LOGic:PODB:BIT<x>:DISPlay?
<x> = 0 to 7
Example
:LOGIC:PODB:BIT1:DISPLAY ON
:LOGIC:PODB:BIT1:DISPLAY?
-> :LOGIC:PODB:BIT1:DISPLAY 1
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODB:BIT<x>:LABel
Function Sets or queries the user-defined threshold level
of a bit of logic input port B.
Syntax
:LOGic:PODB:BIT<x>:LABel {<String>}
:LOGic:PODB:BIT<x>:LABel?
<x> = 0 to 7
<String> = Up to 8 characters
Example
:LOGIC:PODB:BIT1:LABEL "ABC"
:LOGIC:PODB:BIT1:LABEL?
-> :LOGIC:PODB:BIT1:LABEL "ABC"
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODB:BIT<x>:LEVel
Function Sets or queries the label of a bit of logic input port
B.
Syntax
:LOGic:PODB:BIT<x>:LEVel {<Voltage>}
:LOGic:PODB:BIT<x>:LEVel?
<x> = 0 to 7
<Voltage> = −10 to 10 V (in 0.1-V steps)
Example
:LOGIC:PODB:BIT1:LEVEL−10
:LOGIC:PODB:BIT1:LEVEL?
->:LOGIC:PODB:BIT1:LEVEL−10
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODB:BIT<x>:TYPE
Function Selects the threshold level of a bit of logic input
port B.
Syntax
:LOGic:PODB:BIT<x>:TYPE {CMOS1|
CMOS2|CMOS3|CMOS5|ECL}
:LOGic:PODB:BIT<x>:TYPE?
<x> = 0 to 7
Example
:LOGIC:PODB:BIT1:TYPE CMOS1
:LOGIC:PODB:BIT1:TYPE?
-> :LOGIC:PODB:BIT1:TYPE CMOS1
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODB:DESKew
Function Sets or queries the deskewing of logic input port B.
Syntax
:LOGic:PODB:DESKew {<Time>}
:LOGic:PODB:DESKew?
<Time> = −100 to 100 ns (in 10-ps steps)
Example
:LOGIC:PODB:DESKEW−100
:LOGIC:PODB:DESKEW?
->:LOGIC:PODB:DESKEW−100
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODB:HYSTeresis
Function Sets or queries the hysteresis of logic input port B.
Syntax
:LOGic:PODB:HYSTeresis {HIGH|LOW}
:LOGic:PODB:HYSTeresis?
Example
:LOGIC:PODB:HYSTERESIS HIGH
:LOGIC:PODB:HYSTERESIS?
-> :LOGIC:PODB:HYSTERESIS HIGH
Description An error will occur if 16 bit input (/L16 option) is
not supported.
:LOGic:PODL?
Function Queries all settings of logic input port L.
Syntax
:LOGic:PODL?
:LOGic:PODL:ALL?
Function Queries all bit settings of logic input port L.
Syntax
:LOGic:PODL:ALL?
:LOGic:PODL:ALL:DISPlay
Function Turns on or off all bit displays of logic input port L.
Syntax
:LOGic:PODL:ALL:DISPlay {<Boolean>}
Example
:LOGIC:PODL:ALL:DISPLAY ON
:LOGic:PODL:ALL:LEVel
Function Sets or queries the user-defined threshold level
logic input port L.
Syntax
:LOGic:PODL:ALL:LEVel {<Voltage>}
:LOGic:PODL:ALL:LEVel?
<Voltage> = −10 to 10V (in 0.1-V steps)
Example
:LOGIC:PODL:ALL:LEVEL 1
:LOGIC:PODL:ALL:LEVEL?
-> :LOGIC:PODL:ALL:LEVEL 1.000E+00
:LOGic:PODL:ALL:TYPE
Function Sets or queries the threshold level of logic input
port L.
Syntax
:LOGic:PODL:ALL:TYPE {CMOS1|CMOS2|
CMOS3|CMOS5|ECL|USERdefine}
:LOGic:PODL:ALL:TYPE?
Example
:LOGIC:PODL:ALL:TYPE CMOS1
:LOGIC:PODL:ALL:TYPE?
->:LOGIC:PODL:ALL:TYPE CMOS1
5.19 LOGic Group