Page #336 background imageLoading...
Page #336 background image
5-303
IM DLM4038-17EN
Commands
1
2
3
4
5
6
App
Index
:TRIGger[:ATRigger]:PSI5:MODE
Function Sets or queries the PSI5 signal trigger mode.
Syntax
:TRIGger[:ATRigger]:PSI5:MODE {DATA|
SBIT|SYNC}
:TRIGger[:ATRigger]:PSI5:MODE?
Example
:TRIGGER:ATRIGGER:PSI5:MODE DATA
:TRIGGER:ATRIGGER:PSI5:MODE?
-> :TRIGGER:ATRIGGER:PSI5:MODE DATA
:TRIGger[:ATRigger]:PSI5:SETup?
Function Queries all bus setup settings for PSI5 signal
triggering.
Syntax
:TRIGger[:ATRigger]:PSI5:SETup?
:TRIGger[:ATRigger]:PSI5:SETup:DATA?
Function Queries all PSI5 signal trigger data signal
settings.
Syntax
:TRIGger:ATRigger:PSI5:SETup:DATA?
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
BRATe
Function Sets or queries the PSI5 signal trigger bit rate
(data transfer rate).
Syntax
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
BRATe {<NRf>|USER,<NRf>}
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
BRATe?
<NRf> = 125000, 189000
USER <NRf> = See the DLM4000 Features
Guide for this information.
Example
:TRIGGER:ATRIGGER:PSI5:SETUP:DATA:BR
ATE 125000
:TRIGGER:ATRIGGER:PSI5:SETUP:DATA:BR
ATE?
-> :TRIGGER:ATRIGGER:PSI5:SETUP:DATA
:BRATE 125000
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
DBITs
Function Sets or queries the data length for PSI5 signal
triggering.
Syntax
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
DBITs {<NRf>}
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
DBITs?
<NRf> = 10, 16
Example
:TRIGGER:ATRIGGER:PSI5:SETUP:DATA:DB
ITS 10
:TRIGGER:ATRIGGER:PSI5:SETUP:DATA:DB
ITS?
-> :TRIGGER:ATRIGGER:PSI5:SETUP:DATA
:DBITS 10
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
EDETection
Function Sets or queries the error detection method for
PSI5 signal triggering.
Syntax
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
EDETection {CRC|PARity}
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
EDETection?
Example
:TRIGGER:ATRIGGER:PSI5:SETUP:DATA:ED
ETECTION CRC
:TRIGGER:ATRIGGER:PSI5:SETUP:DATA:ED
ETECTION?
-> :TRIGGER:ATRIGGER:PSI5:SETUP:DATA
:EDETECTION CRC
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
SOURce
Function Sets or queries the data source for PSI5 signal
triggering.
Syntax
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
SOURce {<NRf>}
:TRIGger[:ATRigger]:PSI5:SETup:DATA:
SOURce?
<NRf> = 1 to 8
Example
:TRIGGER:ATRIGGER:PSI5:SETUP:DATA:SO
URCE 1
:TRIGGER:ATRIGGER:PSI5:SETUP:DATA:SO
URCE?
-> :TRIGGER:ATRIGGER:PSI5:SETUP:DATA
:SOURCE 1
:TRIGger[:ATRigger]:PSI5:SETup:SYNC
Function Sets or queries the sync signal for PSI5 signal
triggering.
Syntax
:TRIGger[:ATRigger]:PSI5:SETup:SY
NC {<NRf>|NONE}
:TRIGger[:ATRigger]:PSI5:SETup:SYNC?
<NRf> = 1 to 8
Example
:TRIGGER:ATRIGGER:PSI5:SETUP:SYNC 1
:TRIGGER:ATRIGGER:PSI5:SETUP:SYNC?
-> :TRIGGER:ATRIGGER:PSI5:SETUP:SYN
C 1
5.34 TRIGger Group

Table of Contents

Other manuals for YOKOGAWA DLM4038

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA DLM4038 and is the answer not in the manual?

YOKOGAWA DLM4038 Specifications

General IconGeneral
BrandYOKOGAWA
ModelDLM4038
CategoryTest Equipment
LanguageEnglish

Related product manuals