5-316
IM DLM4038-17EN
:WPARameter<x>:WAIT?
Function Waits for the completion of automated
measurement with a set timeout.
Syntax
:WPARameter<x>:WAIT? {<NRf>}
<x> = 1 to 2
<NRf> = 1 to 36000 (timeout, 100 ms)
Example
:WPARAMETER1:WAIT?
-> :WPARAMETER1:WAIT 1
Description The command returns zero if the automated
measurement finishes within the specified
timeout. If automated measurement does not
finish, or if it was never taking place to begin with,
the command returns 1. Even if you set a long
timeout, the command will return zero as soon as
automated measurement finishes.