Appendix 2 Additional Explanation for
Communication Commands
This section explains the units that can be set using communication commands when the pressure unit
suffix code is -U2.
Standard Command
:SENSe:UNIT
Function Sets or queries the pressure unit.
Syntax :SENSe:UNIT {PA|HPA|KPA|MPA|MBAR|BAR|ATM|MMH2O4|MMH2O20|INH2O4|
INH2O20|FTH2O4|FTH2O20|MMHG|INHG|GFCM2|KGFCM2|PSI|TORR}
PA: Pa
HPA: hPa
KPA: kPa
MPA: MPa
MBAR: mbar
BAR: bar
ATM: atm
MMH2O4: mmH
2
O@4°C
MMH2O20: mmH
2
O@20°C
INH2O4: inH
2
O@4°C
INH2O20: inH
2
O@20°C
FTH2O4: ftH
2
O@4°C
FTH2O20: ftH
2
O@20°C
MMHG: mmHg
INHG: inHg
GFCM2: gf/cm
2
KGFCM2: kgf/cm
2
PSI: psi
TORR: Torr
Example :SENSE:UNIT KPA
:SENSE:UNIT? -> :SENSE:UNIT KPA
Compatible Command
PU/PU?
Function Sets or queries the pressure unit.
Syntax PUm<terminator>
PU?<terminator>
m = 1 to 7
1: inH
2
O@4°C
2: inHg
3: psi
4: kPa
5: kgf/cm
2
6: mmH
2
O@4°C
7: mmHg
Example PU4
Description When percentage display or %ERROR display is in use, you cannot set the unit.

Table of Contents

Other manuals for YOKOGAWA MT300

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MT300 and is the answer not in the manual?

YOKOGAWA MT300 Specifications

General IconGeneral
BrandYOKOGAWA
ModelMT300
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals