Appendix 1.1 Commands
Communication Commands 1
• While recalling or storing is in progress,
execution error 19 will occur.
SC/SC? Determines whether or not to use the
scaling function/inquires about the
current setting.
Syntax SC m <terminator>
“m” indicates whether scaling is ON or OFF.
m=0 : OFF
1: ON
Query SC? <terminator>
Example SC1
Description • Parameter error 12 will occur if “m” is set to an
illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
SL Recalls set-up parameters from a
selected file.
Syntax SL m <terminator>
“m” indicates file no., and must be set within the
following range.
1 m 4
Description • Parameter error 12 will occur if “m” is set to an
illegal value.
• It is not possible to recall communications-
related information (communication mode,
address etc.) using this command.
• While recalling or storing is in progress,
execution error 19 will occur.
SO/SO? Sets the store function ON/OFF or
inquires about the current setting.
Syntax SO m <terminator>
“m” indicates whether storage is ON or OFF.
m=0 : OFF
1: ON
Query SO? <terminator>
Example SO1
Description • Parameter error 12 will occur if “m” is set to an
illegal value.
• While recalling or storing is in progress,
execution error 19 will occur.
SR/SR? Sets the storage interval/inquires about
the current setting.
Syntax SR m1,m2,m3 <terminator>
“m1” indicates the hour
0 m1 99
“m2” indicates the minutes
0 m2 59
“m3” indicates the seconds
0 m3 59
Query SR? <terminator>
Example SR0,0,0
Description • Parameter error 12 will occur if an illegal value
is set.
• When the storage interval is set to 0hrs, 0min,
0sec, refer to page 9-2.
• While recalling or storing is in progress,
execution error 19 will occur.
SS Stores set-up parameters into a selected
file.
Syntax SS m <terminator>
“m” indicates file no., and must be set within the
following range.
1 m 4
Description • Parameter error 12 will occur if “m” is set to an
illegal value.
• The following set-up parameters can be stored:
All set-up parameters which can be output by
the OS command
Information related to communications (GP-
IB, RS-232-C etc.)
• While recalling or storing is in progress,
execution error 19 will occur.
TM/TM? Sets integration preset time/inquires
about the current setting.
Syntax TM m1,m2 <terminator>
“m1” indicates hour, and must be set within the
following range.
0 m1 999
“m2” indicates minute, and must be set within the
following range.
0 m2 59
Query TM? <terminator>
Example TM0,0
Description • Parameter error 12 will occur if an illegal value
is set.
• While recalling or storing is in progress,
execution error 19 will occur.
WR/WR? Sets the wiring system/inquires about the
current setting.
Syntax WR m <terminator>
m=2 :1Φ3W
3:3Φ3W
4:3Φ4W (available only for the 253503)
5 :3V3A (available only for the 253503)
Query WR? <terminator>
Example WR2
Description • Parameter error 12 will occur if an illegal value
is set.
YC/YC? Sets the display channel while the
comparator function is ON/inquires about
the current setting.
Syntax YC m <terminator>
“m” indicates the channel number for display
• in case of single mode
m=1 :Displays limit and measurement value
on display 1
2 :Displays limit and measurement value
on display 2
3 :Displays limit and measurement value
on display 3
4 :Displays limit and measurement value
on display 4
• in case of dual mode
m=1,2 :Displays limit and measurement value
on display 1 and 2 respectively
m=3,4 :Displays limit and measurement value
on display 3 and 4 respectively
Query YC? <terminator>
Example YC1
Description • Parameter error 12 will occur if “m” is set to an
illegal value.
YM/YM? Sets the mode of the comparator function/
inquires about the current setting.
Syntax YM m <terminator>
“m” indicates the display mode
m=0 :Single mode
1 :Dual mode
Query YM? <terminator>
Example YM1
Description • Parameter error 12 will occur if “m” is set to an
illegal value.
YO/YO? Sets the comparator function ON/OFF or
inquires about the current setting.
Syntax YO m <terminator>
“m” indicates whether the comparator function is
ON/OFF
m=0 :OFF
1 :ON
Query YO? <terminator>
Example YO1
Description • Parameter error 12 will occur if “m” is set to an
illegal value.

Table of Contents