EasyManuals Logo

YOKOGAWA WT110 User Manual

YOKOGAWA WT110
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #189 background imageLoading...
Page #189 background image
IM 253401-01E
App2-32
Appendix 2.3 Commands
2.3.14 STORe Group
The commands in the STORe group are used to make settings relating to and inquiries about storing data. This allows you to
make the same settings as when using the lower menus of [MEMORY]-”StorE” or [MEMORY]-”PnLSt”.
:STORe : STAT e <Space> OFF
ON
<NRf>
?
INTer val <Space> <NRf> , <NRf> , <NRf>
?
PANel <Space> <NRf>
?
<Character string>
;
STORe?
Function Queries all settings related to storing data.
Syntax
STORe?
Example
STORE?
:STORE:STATE 0;INTERVAL 0,0,0
STORe:INTerval
Function Sets the interval for storage/queries the current setting.
Syntax
STORe:INTerval {<NRf>,<NRf>,<NRf>|<String>}
STORe:INTerval?
{<NRf>,<NRf>,<NRf>}=0,0,0 to 99,59,59
{<String>}=HH:MM:SS HH
hour
MM
min
SS
sec
Example
STORE:INTERVAL 0,0,0
STORE:INTERVAL “00:00:00“
STORE:INTERVAL?
:STORE:INTERVAL 0,0,0
Description • If the storage interval is set to 0,0,0, the storage
interval becomes 250ms in case of normal
measurement.
• For the storage interval in case of harmonic analysis,
refer to page 9-2.
STORe:PANel
Function Saves the set-up parameters to a file.
Syntax
STORe:PANel {<NRf>}
{<NRf>}=1 to 4:file number
Example
STORE:PANEL 1
STORe[:STATe]
Function Sets store ON/OFF, queries the current setting.
Syntax
STORe[:STATe] {<Boolean>}
STORe:STATe?
Example
STORE:STATE ON
STORE:STATE?
:STORE:STATE 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA WT110 and is the answer not in the manual?

YOKOGAWA WT110 Specifications

General IconGeneral
BrandYOKOGAWA
ModelWT110
CategoryMeasuring Instruments