EasyManua.ls Logo

York YLUA0078 - Table 37 - PRESSURE TRANSDUCERS; System 1 Discharge Transducer; System 2 Discharge Transducer; System 1 Suction Transducer

York YLUA0078
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
JOHNSON CONTROLS
147
SECTION 9 – SERVICE AND TROUBLESHOOTING
FORM 150.73-NM1
ISSUE DATE: 09/04/2020
9
TABLE 37 - PRESSURE TRANSDUCERS
0-400 PSIG SUCTION
PRESSURE TRANS-
DUCER
0-600 PSIG DISCHARGE
PRESSURE TRANS-
DUCER
PRESSURE
PSIG
VOLTAGE
VDC
PRESSURE
PSIG
VOLTAGE
VDC
0 0.5 0 0.5
50 1.0 75 1.0
100 1.5 150 1.5
150 2.0 225 2.0
200 2.5 300 2.5
250 3.0 375 3.0
300 3.5 450 3.5
350 4.0 525 4.0
400 4.5 600 4.5
Red Wire = 5V, Black wire = 0V, White/Green Wire = signal
TEST POINTS:
Suction Pressure:
System 1:
....................................
Microboard J7-10 to J7-9
System 2:
....................................
Microboard J9-10 to J9-9
DISCHARGE PRESSURE:
System 1:
....................................
Microboard J7-11 to J7-7
System 2:
....................................
Microboard J9-11 to J9-7
V = (Pressure in psig x .01) + .5
or
V = (Pressure in BARG x .145) + .5
where V = dc voltage output
Pressure = pressure sensed by transducer
The I/O board connections for the Discharge Transduc-
ers:
System 1 Discharge Transducer
J7-6 = +5 VDC regulated supply to transducer.
J7-11 = VDC input signal to the microboard.
See the formula above for voltage readings that
correspond to specic discharge pressures.
J7-7 = +5 VDC return
J7-2 = drain (shield connection = 0 VDC)
System 2 Discharge Transducer
J9-6 = +5 VDC regulated supply to transducer.
J9-11 = VDC input signal to the microboard.
See the formula above for voltage readings that
correspond to specic discharge pressures.
J9-7 = +5 VDC return
J9-2 = drain (shield connection = 0 VDC)
The suction transducers have a range from 0 psig to
400 psig (27.5 barg). The output will be linear from 0.5
VDC to 4.5 VDC over the 400 psig (27.5 barg) range.
Following is a formula that can be used to verify the
voltage output of the transducer. All voltage reading
are in reference to ground (unit case).
V = (Pressure in psig x .02) + .5
or
V = (Pressure in barg x .29) + .5
where V = dc voltage input to micro
Pressure = pressure sensed by transducer
Following are the I/O board connections for the Suc-
tion Transducer:
System 1 Suction Transducer
J7-5 = +5 VDC regulated supply to transducer.
J7-10 = VDC input signal to the microboard.
See the formula above for voltage readings that
correspond to specic suction pressures.
J7-9 = +5 VDC return
J7-1 = drain (shield connection = 0VDC)
System 2 Suction Transducer
J9-5 = +5 VDC regulated supply to transducer.
J9-10 = VDC input signal to the microboard.
See the formula above for voltage readings that
correspond to specic suction pressures.
J7-9 = +5 VDC return
J7-11 = drain (shield connection = 0 VDC)

Table of Contents

Related product manuals