Section 6.0
MAN 06–0017H
ZIEHM QUANTUM User’s Manual Page 80 of 220
Control No. Symbol(s) Description
B
utton 32
Metal
Activates the metal artifact correction
function.
This is a supplementary operation to the
organ program image quality.
B
utton 33
Soft
The Soft button has been reprogrammed to
penetrate larger patient anatomy ( Adipose
patients) and is called “LPD” on the Desk
View touch screen
See also Button 29
B
utton 34
Not Active
Window/Level
Will not activate a function, this button is
blocked
B
utton 35
Tube kV value
Shows the automatically determined or
manually set tube voltage in kV
After Fluoroscopy, the last kV value remains
stored and displayed.
B
utton 36
Tube Current mA
Shows the automatically determined tube
current in mA.
After Fluoroscopy, the last value remains
stored and displayed.
B
utton 37
Radiography Time
Displays the radiography exposure time in
seconds