EasyManua.ls Logo

Zmorph i500 - Nozzle Calibration

Zmorph i500
69 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Product Manual - ZMORPH i500 back to table of content
>>
40
Maintenance
3. Once done connect the USB ash drive (included with the machine) to the front USB slot. Before the
operation please make sure the bed is empty and the nozzle is clean.
4. From main menu go to the “choose le” tab and select “USB memory. Go to the Calibration folder
and select the “XY calibration.gcode” le. At next start the gcode (the XY calibration procedure will start).
Please note, the XY calibration require both laments loaded.
1. The calibration begin with
providing the X and Y oset into
the machine. From main menu go
to the CALIBRATION menu.
2. select “XY Oset”. Set X
oset at -20 and Y oset to 0,
next press the Set X and Set Y
buttons, once done press Apply
Osets.
10.4.2 Nozzle Calibration
Nozzle Calibration allows to set the oset of the nozzles in X and Y axes.

Related product manuals