EasyManua.ls Logo

ABB Advance Optima Uras 14 - Optical Alignment Procedures

ABB Advance Optima Uras 14
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SHB_U14 Chapter 8: Calibration
8-3
Optical Alignment
Accessories Screwdriver
Emitter wrench, part number 096 79 80
Zero gas
PC with test and calibration software
(after AO software version 1.3 can also be performed via the AO user interface)
When should optical
alignment be
performed?
After removing/installing beam path components
Emitter, sample cell, calibration cell, optical filter, filter cell, IR detector
Offset value greater than 20%
After repairing sample gas path leaks (zero gas distortion)
After cleaning the sample cell
Routinely every 6 to 12 months (depends on operating conditions)
Note Optical alignment causes the emitter intensity to be equal on the sample and
reference sides of the IR detector.
Each beam path in an analyzer module must be optically aligned separately.
If there are two IR detectors in a beam path, select the rear detector for
alignment.
F
No optical alignment is possible for an oxygen sensor.
Preparation Establish communication with the analyzer module via TCT.
Menu: Calibration
Connect the zero gas
- Analyzer module with sample gas/zero gas solenoid valve:
Connect the zero gas to the module zero gas port.
- Analyzer module without solenoids:
Connect the zero gas to the module sample gas port.
Continued on next page

Table of Contents

Related product manuals