Loading...
5-12 Chapter 5: Testing SHB_U14
Thermal link
Accessories
Multimeter, screwdriver
Electrical value:
Resistance
Where?
Temperature control board terminal strip X2
Remove the thermal link from the terminal strip and measure the thermal link
continuity.
Desired value:
Continuity
Figure 5-10
Thermal link