7-46 Chapter 7: Configuration SHB_U14
Solenoids
General
To calibrate the analyzer module from one to three optional solenoids can be
installed and then set up using the Gas module menu item.
1 solenoid: For analyzer modules with calibration cell and possible
additional oxygen sensor.
SAMPLE GAS SOLENOID 3/2-way
3 solenoids: For analyzer modules without calibration cells, but with internal
calibration.
SAMPLE GAS SOLENOID 3/2-way
ZERO GAS SOLENOID 2/2-way
END GAS SOLENOID 2/2-way
Select SAMPLE GAS SOLENOID if only one solenoid is set up.
Solenoid valve
configuration
Continued on next page