8-24 Chapter 8: Calibration
SHB_U14
Pressure detector initial calibration
When should an
initial calibration be
performed?
If a pressure correction is made
After changing the pressure sensor
Accessories Pressure detector calibration can be performed using the keypad on the
device or with a PC and the test and calibration software. Both methods are
based on the same calculation and evaluation.
For calibration with a PC:
PC with serial port
Advance Optima - TCT test and calibration software
Serial port connecting cable, item no. 0743091
Pressure gauge for ambient air pressure
Preparation Establish electrical connections
The analyzer should be installed at the site at which it is to be used and should
be protected against vibration
Set the test gas flows between 30 and 60 l/h and maintain stable +/-5 l/h.
Connect the PC to the analyzer module
Start the calibration and test software
Test Test all detectors (the measured values should appear and be regularly
updated)
Status message 0 for all detectors
Continued on next page