SHB_U14 Chapter 5: Testing 5-9
Thermostat temperature
Accessories
AO service unit, PC with MASSERV
Where? • AO service unit
Select using the Diagnostics/Info - Module-specific - Controller parameter menu
• PC with MASSERV
• Desired value:
Menu F1 Configure - F2 Detector configuration - Temperature detector control 2
- F5 Detector components - F3 Component correction parameter - Controller
parameter
• Actual value:
F4 Device test menu
Desired value
Temperature: 55°C
Raw measured value: approx. 2000
Controller variable: Depends on ambient temperature
approx. 40% at 25°C
approx. 10-20% at 40°C
Limit 1:
±2°C
Limit 2:
±10°C