5-14 Chapter 5: Testing SHB_U14
Calibration unit
Accessories
Multimeter, PC with test- and calibration software
Visual check
The upper, lower and lateral travel paths of the calibration cell slide must be free of
interference by cables and tubes.
Function check • The calibration cell installed in the calibration unit should travel to the lower
position when the analyzer module is turned on (measurement/zero calibration).
• The calibration cells can be moved individually using MASSERV.
Device test menu
Electrical value:
Voltage
Where?
Plugs X6 or X7 Pin 2 + to pin 1
⊥
Desired value:
Power supply: +15V
Figure 5-11
Connection cable 1