5-18 Chapter 5: Testing SHB_U14
Sensor-CPU circuit board
Accessories
Multimeter
Preface
Always replace a defective Sensor-CPU board.
The following test can be performed to determine which function is defective.
Visual check LEDs
H1 green 24 V power supply OK
H2 yellow Thermostat temperature still outside limits
H3 red Defective Sensor-CPU circuit board
Normal condition: LED lights shortly after the analyzer module is
turned on.
SMD fuses
F1 0.2 A 24 V power for solenoid 1 "......."
F2 3.15 A 24 V power for thermostat heater
F3 3.15 A 24 V power for internal pump
F4 0.2 A 24 V power for solenoid 2 "......."
F5 0.2 A 24 V power for solenoid 3 "......."
Electrical value
Voltage
Where? Sensor-CPU circuit board, plugs X8 and X10
Desired value
Function Plug Pin Desired value
Off On
Heater X 8 1 2 0 V 24 V intermitt.
Pump X10 2 3 0 V 0-24 V
Solenoid 1 X10 5 6 0 V 24 V
Solenoid 2 X10 7 8 0 V 24 V
Solenoid 3 X10 9 10 0 V 24 V
Flow detector
Internal gas lines
X10 17 20 ()
18 20 ()
+ 15 V
- 15 V
Fortsetzung auf der folgenden Seite