SHB_U14 Chapter 8: Calibration
8-19
Delta phase
When should a Delta
phase be adjusted?
As a rule it is done for CO, NO and SO
2
in small measurement ranges
The cross sensitivity effect of some interference components can be fine-tuned for
these sample components by adjusting a Delta phase.
This cross sensitivity alignment is only performed as needed, i.e. if the applicable
associated components are present.
In general, in flue gas applications compensation is made for 3°C H
2
O and 15% Vol
CO
2
interference components.
Accessories • PC with serial port
• Advance Optima - TCT test and calibration software
• Serial port connecting cable, item no. 0743091
• Test gases
• Flow measurement unit
Requirements Automatic phase alignment performed
Initial calibration performed
Continued on next page