SHB_U14 Chapter 6: Component Interchange 6-3
Remove/install Uras 14 analyzer module
Module removal
To remove the module, proceed as follows:
Step Action
1 Turn off the analyzer power supply.
2 Shut off the gas supply (sample gas and, if applicable, reference gas)
to the analyzer.
3 Disconnect the gas lines from the analyzer module ports.
4 Flush the analyzer module.
5 Open the system housing.
6 Remove the cables connecting the analyzer module to the central unit.
7 Disconnect the hood cable.
8 Remove 4 screws (1)
9 Remove the hood.
10 Remove the analyzer module mounting screws.
4 bolts on the gas connection plate (2).
4 nuts on the base plate (3).
11 Remove the analyzer module from the housing
Module installation
To install the module, proceed as follows
Step Action
1 Place the analyzer module in the system housing.
2 Secure the analyzer module with the four bolts and four nuts.
3 Install the hood.
4 Reconnect the cables.
5 Close the system housing.
6 Check the analyzer module seal integrity.
7 Connect the gas lines to the analyzer module.
8 Open the gas supply to the analyzer module.
9 Turn on the analyzer system power supply
Continued on the following page