5-16 Chapter 5: Testing SHB_U14
IR module circuit board
Accessories
Multimeter
Preface
Always replace a defective IR module board.
The following test can be performed to determine which function is defective.
Visual check
SMD fuses
F1 24 V power supply emitter 1
F2 24 V power supply emitter 2
Electrical value
Voltage
Where? ÌR module circuit board, plugs X3 and X4
Desired value
Function Plug Pin Desired value
IR detector 1 X3 20 19 ()
24 25 ()
28 25 ()
+150 V
+ 15 V
- 15 V
IR detector 2 X3 30 29 ()
34 35 ()
38 35 ()
+ 150 V
+ 15 V
- 15 V
IR detector 3 X4 2 1 ()
6 7 ()
10 7 ()
+ 150 V
+ 15 V
- 15 V
IR detector 4 X4 12 11 ()
16 17 ()
17 20()
+ 150 V
+ 15 V
- 15 V
Motor control X3
Beam splitter X3 13 12 () + 5 V
Pressure sensor X3 15 17 ()
16 17 ()
+ 15 V
- 15 V
Calibration cell control 1 X3 41 42() + 15 V
Calibration cell control 2 X3 47 48() + 15 V
Emitter 1 X3 3 5 (-) + 5...10 V
Emitter 2 X3 7 6 (-) + 5... 10V
Continued on next page