SHB_U14 Chapter 8: Calibration
8-17
Cross-sensitivity alignment
When should a cross-
sensitivity alignment
be performed?
Perform a CS correction if a sample component is inadequately affected by another
component (>2% of sample component span).
If the interference components are to be measured with the same unit as the
sample component, perform an internal CS alignment.
For instance, CS corrections are required in the following applications:
Sample component N
2
O, interference component CO
2
All required CS corrections can be found in the "Rekonur" configuration program.
Accessories PC with serial port
Advance Optima - TCT test and calibration software
Serial port connecting cable, item no. 0743091
Test gases
Flow measurement unit
Preparation Establish electrical connections
The analyzer should be installed at the site at which it is to be used and should
be protected against vibration
Set the test gas flows between 30 and 60 l/h and maintain stable +/-5 l/h.
Connect the PC to the analyzer module
Start the calibration and test software
Test Test all detectors (the measured values should appear and be regularly
updated)
Status message 0 for all detectors
Continued on next page