EasyManua.ls Logo

ABB Advance Optima Uras 14 - Postlinearization Procedures

ABB Advance Optima Uras 14
200 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-14 Chapter 8: Calibration
SHB_U14
Postlinearization
Accessories Zero gas, span gas
Mix pump or applicable test gas
PC with TCT
When should
postlinearization be
performed?
After an IR detector change
In case of excessive linearity deviation (> 1% of the applicable measurement
range)
After reconfiguring for other measurement ranges
Note Optimum postlinearization results are obtained with a test gas and mix pump.
Since test gases normally have a tolerance of up to 2%, postlinearization with a
test gas will be less precise than postlinearization by mixing with a pump.
The characteristics of each sample component are determined by parameters a
1
to a
4
. By adapting the two postlinearization parameters, postlinearization provides
for an individual linearity setting.
An initial calibration does not affect these two parameters;
therefore, if postlinearization is impossible after a detector change due to the lack
of a mix pump, these two parameters should be set to 1.
A calibration reset does not affect these two parameters.
If the large measurement range < 0.15 < cl
max
, the postlinearization will calculate
a new linearity function with parameters a
1
und a
2
. (this occurs with very small
measurement ranges)
Parameters a
3
and a
4
are then 0.
The two postlinearization parameters are 1.
Preparation The IR detectors must be calibrated or have undergone an initial calibration.
Establish communication with the analyzer module via TCT.
Connect zero or span gas to the sample gas port.
Set the flow rate to the 30-60 l/h range:
Continued on next page

Table of Contents

Related product manuals